Specifi cations, continued – Atec Agilent-6060B User Manual
Page 4
![background image](/manuals/318508/4/background.png)
4
Specifi cations, continued
Supplemental characteristics
(Non-warranted characteristics determined
by design and useful in applying the product)
6060B
6063B
Analog monitor accuracy
Current monitor (0 to 10 V
out
)
4% ± 85 mA
3% ± 10 mA
Temperature coeffi cient
50 ppm/°C ± 6 mA/°C
100 ppm/°C ± 1 mA/°C
Voltage monitor (0 to 10 V
out
)
0.25% ± 40 mV
0.4% ± 240 mV
Temperature coeffi cient
50 ppm/°C ± 0.2 mV/°C
70 ppm/°C ± 1.2 mV/°C
Remote sensing
5-VDC maximum between sense and
load input
5-VDC maximum between sense and
load input
Minimum operating voltage
(at full rated current)
2 V (1.2 V typical)
2 V (1.2 V typical)
Programmable short
0.033 Ω (0.020 Ω typical)
0.20 Ω (0.10 Ω typical)
Programmable open (typical)
20 kΩ
80 kΩ
Drift (over 8-hour interval)
Current
0.03% ± 10 mA
0.03% ± 15 mA
Voltage
0.01% ± 10 mV
0.01% ± 20 mV
DC isolation voltage
± 240 VDC, between any input and
chassis ground
± 240 VDC, between any input and
chassis ground
Digital inputs
V
IL
= 0.9 V
max
at I
IL
=
–1 mA / V
IH
= 3.15 V
min
(pull-up resistor on input)
V
IL
= 0.9 V
max
at I
IL
=
–1 mA / V
IH
= 3.15 V
min
(pull-up resistor on input)
Digital outputs
V
OL
= 0.72 V
max
at I
OL
=
1 mA / V
OH
= 4.4 V
min
at I
OH
= –20 μA
V
OL
= 0.72 V
max
at I
OL
=
1 mA / V
OH
= 4.4 V
min
at I
OH
= –20 μA
Notes:
1. Operating temperature range is 0° to 55 °C. All specifi cations apply for 25 °C ± 5 °C, except as noted.
2. Maximum continuous power available is derated linearly from 40 °C to 75% of maximum at 55 °C.
3. DC current accuracy specifi cations apply 30 seconds after input is applied.