Specifi cations
Specifications
(at 25 °C ± 5 °C unless otherwise specified)
6060B
6063B
DC Input ratings
Current
0 to 60 A
0 to 10 A
Voltage
3 to 60 V
3 to 240 V
Maximum power (at 40° C)
300 W
250 W
Constant current mode
Ranges
0 to 6 A, 0 to 60 A
0 to 1 A, 0 to 10 A
Accuracy
0.1% ± 75 mA
0.15% ± 10 mA
Regulation
10 mA
8 mA
Constant voltage mode
Accuracy
0.1% ± 50 mV
0.12% ± 120 mV
Regulation (w/remote sense)
10 mV
10 mV
Constant resistance mode
Ranges
0.033 to 1.0 Ω
0.20 to 24.0 Ω
1 to 1,000 Ω
24 to 10,000 Ω
10 to 10,000 Ω
240 to 50,000 Ω
Accuracy
1 Ω: 0.8% ± 8 mΩ (with ≥ 6 A at input)
24 Ω: 0.8% ± 200 mΩ (with ≥ 1 A at input)
1 KΩ: 0.3% ± 8 mΩ (with ≥ 6 V at input)
10 KΩ: 0.3% ± 0.3 mΩ (with ≥ 24 V at input)
10 KΩ: 0.3% ± 8 mΩ (with ≥ 6 V at input)
50 KΩ: 0.3% ± 0.3 mΩ (with ≥ 24 V at input)
Transient generator
Frequency range
0.25 Hz to 10 kHz
0.25 Hz to 10 kHz
Accuracy
3%
3%
Duty cycle range
3 to 97% (0.25 Hz to 1 kHz)
3 to 97% (0.25 Hz to 1 kHz)
6 to 94% (1 to 10 kHz)
6 to 94% (1 to 10 kHz)
Accuracy
6% of setting ± 2%
6% of setting ± 2%
Current level high range
0 to 60 A
0 to 10 A
Accuracy
0.1% ± 350 mA
0.18% ± 50 mA
Current level low range
0 to 6 A
0 to 1 A
Accuracy
0.1% ± 80 mA
0.18% ± 13 mA
Voltage level
0 to 60 V
0 to 240 V
Voltage level accuracy
0.1% ± 300 mV
0.15% ± 1.1 V
Readback accuracy
Current
0.05% ± 65 mA
0.12% ±10 mA
Voltage
± (0.05% + 45 mV)
± (0.1% + 150 mV)
Ripple and noise (20 Hz to 10 MHz noise)
Current
4 mA rms
1 mA rms
40 mA peak-to-peak
10 mA peak-to-peak
Voltage
6 mV rms
6 mV rms