Dc power analyzer mainframe key characteristics – Atec Agilent-N6700 Series User Manual
Page 13
13
DC Power Analyzer Mainframe Key Characteristics
Interface capabilities
GPIB
SCPI – 1993, IEEE 488.2 compliant interface
LXI compliance
Class C
USB 2.0
Requires Agilent IO Library version M.01.01 or 14.0 and up
10/100 LAN
Requires Agilent IO Library version L.01.01 or 14.0 and up
Environmental conditions
Operating environment
Indoor use, installation category II (for AC input), pollution degree 2
Temperature range
0 °C to 55 °C (output current is derated 1% per °C above 40 °C ambient temperature)
Relative humidity
Up to 95%
Altitude
Up to 2000 meters
AC input
Input ratings
~ 100 VAC – 240 VAC; 50/60/400 Hz
Power consumption
1000 VA (N6700B)
1440 VA (N6701A)
1440 VA (N6702A @ < 180 VAC input)
2200 VA (N6702A @ > 180 VAC input)
Power factor
0.99 @ nominal input and rated power
Net weight
N6700B with 4 modules (typical)
12.73 kg/28 lbs.
N6701A with 4 power modules
11.82 kg/26 lbs.
N6702A with 4 power modules
14.09 kg/31 lbs.
Dimensions
N6700B /N6701A
N6702A
Height/width/depth
44.45 mm/432.5 mm/585.6 mm
1.75 in./17.03 in./23.06 in.
44.45 mm/432.5 mm/633.9 mm
1.75 in./17.03 in./24.96 in.