Performance characteristics

(continued)

Trigger system (continued)

Nth edge burst

Trigger on the Nth edge of a burst that occurs after an idle time that you specify. Max edge

count: 65,536.

Autoscale

Finds and displays all active scope and logic (for 7000B Series MSO) channels, sets edge

trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and

thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage

>10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.

Scope channel triggering

Range (internal)

±6 div from center screen

Sensitivity*

<10 mV/div: greater of 1 div or 5 mV; ≥10 mV/div: 0.6 div

Coupling

AC, ~10 Hz on MSO/DSO701xB/703xB/705xB/710xB,

DC, noise reject, HF reject and LF reject (~50 kHz)

Digital (D15 - D0) channel triggering (7000B Series MSO or MSO-upgraded 7000B Series DSO)

Threshold range (user defined)

±8.0 V in 10 mV increments

Threshold accuracy

±(100 mV + 3% of threshold setting)

Predefined thresholds

TTL = 1.4 V, CMOS = 2.5 V, ECL = –1.3 V

External (EXT) triggering

MSO/DSO7xx2B (2-/2+16-ch models)

MSO/DSO7xx4B (4-/4+16-ch models)

Input impedance

1 MΩ ± 3% || 14 pF or 50 Ω

2.14 kΩ ±5%

Maximum input

CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk

±15 V

With 10073C/D 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk

5 Vrms with 50-Ω input

Range

DC coupling: trigger level ±1 V and ±8 V

±5 V

Sensitivity

For ±1 V range setting:

DC to 500 MHz: 500 mV

DC to 100 MHz, 100 mV;

For ±8 V range setting:

DC to 100 MHz, 250 mV;

>100 MHz to bandwidth of oscilloscope: 500 mV

Coupling

AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)

Probe ID

Auto probe sense and AutoProbe interface

Agilent- and Tektronix-compatible passive probe sense

* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

21