Performance characteristics, Continued) – Atec Agilent-7000B Series User Manual
Page 24
Performance characteristics
(continued)
General characteristics (continued)
Trigger out
When Triggers is selected (delay ~17 ns)
0 to 5 V into high impedance
0 to 2.5 V into 50 Ω
When Source Frequency or Source Frequency/8 is selected
0 to 580 mV into high impedance
0 to 290 mV into 50 Ω
Max frequency output: 350 MHz (in source frequency mode when terminated in 50 Ω)
125 MHz (in source frequency/8 mode when terminated in 50 Ω)
10 MHz ref in/out
TTL out, 180 mV to 1 V amplitude with 0 to 2 V offset
Kensington lock
Connection on rear panel for security
Power requirements
Line voltage range
100-120V, 50/60/400 Hz; 100-240V, 50/60 Hz auto ranging
Line frequency
50/60 Hz, 100-240 VAC; 400 Hz, 100-120 VAC
Power usage
120 W max
Environmental characteristics
Ambient temperature
Operating -10 °C to +55 °C; non-operating –40 °C to +70 °C
Humidity
Operating: Up to 95% relative humidity (non-condensing) at +40 °C
Non-operating: Up to 90% relative humidity at +65 °C
Altitude
Operating to 4,570 m (15,000 ft); non-operating to 15,244 m (50,000 ft)
Vibration
Agilent class B1 and MIL-PRF-28800F; class 3 random
Shock
Agilent class B1 and MIL-PRF-28800F; class 3 random; (operating 30g, 1/2 sine,
11 ms duration, 3 shocks/axis along major axis, total of 18 shocks)
Pollution degree
Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation must be expected.
Typical operator noise
30 dBa at front of instrument, 35 dBa at rear of instrument.
Indoor use
Rated for indoor use only
Other
Measurement categories
CAT I
Regulatory information
Safety IEC 61010-1:2001 / EN 61010-1:2001
Canada: CSA C22.2 No. 1010.1:1992
UL 61010B-1:2003
Supplementary information
The product herewith complies with the requirements of the Low Voltage Directive
73/23/EEC and the EMC Directive 89/336/EEC, and carries the CE-marking accordingly.
The product was tested in a typical configuration with HP/Agilent test systems.
24