Performance characteristics

(continued)

Vertical system: scope channels

Scope channels

MSO/DSO7xx2B: Ch 1 and 2 simultaneous acquisition

MSO/DSO7xx4B: Ch 1, 2, 3 and 4 simultaneous acquisition

Bandwidth (–3 dB)*

MSO/DSO701xB: DC to 100 MHz

MSO/DSO703xB: DC to 350 MHz

MSO/DSO705xB: DC to 500 MHz

MSO/DSO710xB: DC to 1 GHz

AC coupled

MSO/DSO701xB: 3.5 Hz to 100 MHz

MSO/DSO703xB: 3.5 Hz to 350 MHz

MSO/DSO705xB: 3.5 Hz to 500 MHz

MSO/DSO710xB: 3.5 Hz to 1 GHz

Calculated rise time (=0.35/bandwidth)

MSO/DSO701xB: 3.5 nsec

MSO/DSO703xB: 1 nsec

MSO/DSO705xB: 700 psec

MSO/DSO710xB: 350 psec

Single-shot bandwidth

MSO/DSO701xB: 100 MHz

MSO/DSO703xB: 350 MHz

MSO/DSO705xB: 500 MHz

MSO/DSO710xB: 1 GHz (in half-channel mode)

Range

1

MSO/DSO701xB, MSO/DSO703xB and MSO/DSO705xB: 2 mV/div to 5 V/div (1 MΩ or 50 Ω)

MSO/DSO710xB: 2 mV/div to 5 V/div (1 MΩ), 2 mV/div to 1 V/div (50 Ω)

Maximum input

CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk

With 10073C/D 10:1 probe: CAT I 500 Vpk

Offset range

±5 V on ranges <10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div;

±75 V on ranges >200 mV/div

Dynamic range

±8 div

Input impedance

1 MΩ ± 1% || 14 pF or 50 Ω ± 1.5%, selectable

Coupling

AC, DC

BW limit

25 MHz selectable

Channel-to-channel isolation

DC to max bandwidth >40 dB

Standard probes

10073D or 10074D shipped standard for each scope channel (N2873A or N2871A optional)

Probe ID

Auto probe sense and AutoProbe interface

Agilent- and Tektronix-compatible passive probe sense

* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

1 2 mV/div is a magnification of 4 mV/div setting for 350 MHz to 1 GHz models. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting

and 32 mV for 2 mV/div sensitivity setting.

17