Performance characteristics
(continued)
Vertical system: scope channels
Scope channels
MSO/DSO7xx2B: Ch 1 and 2 simultaneous acquisition
MSO/DSO7xx4B: Ch 1, 2, 3 and 4 simultaneous acquisition
Bandwidth (–3 dB)*
MSO/DSO701xB: DC to 100 MHz
MSO/DSO703xB: DC to 350 MHz
MSO/DSO705xB: DC to 500 MHz
MSO/DSO710xB: DC to 1 GHz
AC coupled
MSO/DSO701xB: 3.5 Hz to 100 MHz
MSO/DSO703xB: 3.5 Hz to 350 MHz
MSO/DSO705xB: 3.5 Hz to 500 MHz
MSO/DSO710xB: 3.5 Hz to 1 GHz
Calculated rise time (=0.35/bandwidth)
MSO/DSO701xB: 3.5 nsec
MSO/DSO703xB: 1 nsec
MSO/DSO705xB: 700 psec
MSO/DSO710xB: 350 psec
Single-shot bandwidth
MSO/DSO701xB: 100 MHz
MSO/DSO703xB: 350 MHz
MSO/DSO705xB: 500 MHz
MSO/DSO710xB: 1 GHz (in half-channel mode)
Range
1
MSO/DSO701xB, MSO/DSO703xB and MSO/DSO705xB: 2 mV/div to 5 V/div (1 MΩ or 50 Ω)
MSO/DSO710xB: 2 mV/div to 5 V/div (1 MΩ), 2 mV/div to 1 V/div (50 Ω)
Maximum input
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
With 10073C/D 10:1 probe: CAT I 500 Vpk
Offset range
±5 V on ranges <10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div;
±75 V on ranges >200 mV/div
Dynamic range
±8 div
Input impedance
1 MΩ ± 1% || 14 pF or 50 Ω ± 1.5%, selectable
Coupling
AC, DC
BW limit
25 MHz selectable
Channel-to-channel isolation
DC to max bandwidth >40 dB
Standard probes
10073D or 10074D shipped standard for each scope channel (N2873A or N2871A optional)
Probe ID
Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
1 2 mV/div is a magnification of 4 mV/div setting for 350 MHz to 1 GHz models. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting
and 32 mV for 2 mV/div sensitivity setting.
17