Atec Agilent-E5100A User Manual
Page 5
5
• E5100B
without Option E5100B-801/802/803 (at +5 dBm output level, RF attenuator: 0 dB, 50
Ω
input)
10 kHz
≤
freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –85 dB (SPC)
100 kHz
≤
freq. < 250 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –100 dB (SPC)
250 MHz
≤
freq.
≤
300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –95 dB (SPC)
with Option E5100B-801/802/803 (at +16 dBm output level, RF attenuator: 0 dB, 50
Ω
input)
10 kHz
≤
freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –95 dB (SPC)
100 kHz
≤
freq. < 250 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –110 dB (SPC)
250 MHz
≤
freq.
≤
300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –105 dB (SPC)
Residual response (RF attenuator: 0 dB, except for the following points) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –80 dBm
50 kHz, 100 kHz, 95.825 MHz, 95.875 MHz,159.791667 MHz, 159.825 MHz,159.841667 MHz, 159.875 MHz,
239.75 MHz, and 239.875 MHz
Input connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BNC female
with Option E5100x-705/706/707/708 . . . . . . . . . . . . . . . . . . . . . . . . . BNC female, Type-N female (for A, B inputs)
Measurement mode
with Option E5100x-100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A
with Option E5100x-200 or E5100x-600 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A/R, R/A, R, A
with Option E5100x-300 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A/R, B/R, R/A, B/A, R/B, A/B, R, A, B
with Option E5100x-400 . . . . . . . . . . . . . . A/R, B/R, C/R, R/A, B/A, C/A, R/B, A/B, C/B, R/C, A/C, B/C, R, A, B, C
(When the measurement mode is either R/A, B/A, C/A, R/B, A/B, C/B, R/C, or A/C, the specification is SPC.)
Magnitude characteristics
Absolute characteristics
Absolute amplitude accuracy
(at 23 ±5°C, –30 dBm input level for RF attenuator: 0 dB or –5 dBm input level for RF attenuator: 25 dB,
50
Ω
input)
E5100A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
E5100B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±2 dB (SPC)
Ratio characteristics
Frequency response
(at 23 ±5°C, –30 dBm input level for RF attenuator: 0 dB or –5 dBm input level for RF attenuator: 25 dB,
the same RF attenuator setting for both inputs)
50
Ω
input
•E5100A
10 kHz
≤
freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
100 kHz
≤
freq.
≤
100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.5 dB
100 MHz < freq.
≤
300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
• E5100B
100 kHz
≤
freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±2 dB (SPC)
100 kHz
≤
freq.
≤
100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB (SPC)
100 MHz < freq.
≤
300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±2 dB (SPC)
1 M
Ω
input for Option 703, 704, 707, or 708
(using 50
Ω
feedthrough) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±3 dB
Note: Frequency response can be improved by calibration.