beautypg.com

Atec Agilent-E5100A User Manual

Page 5

background image

5

• E5100B

without Option E5100B-801/802/803 (at +5 dBm output level, RF attenuator: 0 dB, 50

input)

10 kHz

freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –85 dB (SPC)

100 kHz

freq. < 250 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –100 dB (SPC)

250 MHz

freq.

300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –95 dB (SPC)

with Option E5100B-801/802/803 (at +16 dBm output level, RF attenuator: 0 dB, 50

input)

10 kHz

freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –95 dB (SPC)

100 kHz

freq. < 250 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –110 dB (SPC)

250 MHz

freq.

300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –105 dB (SPC)

Residual response (RF attenuator: 0 dB, except for the following points) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –80 dBm

50 kHz, 100 kHz, 95.825 MHz, 95.875 MHz,159.791667 MHz, 159.825 MHz,159.841667 MHz, 159.875 MHz,
239.75 MHz, and 239.875 MHz

Input connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BNC female

with Option E5100x-705/706/707/708 . . . . . . . . . . . . . . . . . . . . . . . . . BNC female, Type-N female (for A, B inputs)

Measurement mode

with Option E5100x-100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A
with Option E5100x-200 or E5100x-600 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A/R, R/A, R, A
with Option E5100x-300 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A/R, B/R, R/A, B/A, R/B, A/B, R, A, B
with Option E5100x-400 . . . . . . . . . . . . . . A/R, B/R, C/R, R/A, B/A, C/A, R/B, A/B, C/B, R/C, A/C, B/C, R, A, B, C

(When the measurement mode is either R/A, B/A, C/A, R/B, A/B, C/B, R/C, or A/C, the specification is SPC.)

Magnitude characteristics

Absolute characteristics

Absolute amplitude accuracy
(at 23 ±5°C, –30 dBm input level for RF attenuator: 0 dB or –5 dBm input level for RF attenuator: 25 dB,
50

input)

E5100A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
E5100B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±2 dB (SPC)

Ratio characteristics

Frequency response
(at 23 ±5°C, –30 dBm input level for RF attenuator: 0 dB or –5 dBm input level for RF attenuator: 25 dB,
the same RF attenuator setting for both inputs)

50

input

•E5100A

10 kHz

freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB

100 kHz

freq.

100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0.5 dB

100 MHz < freq.

300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB

• E5100B

100 kHz

freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±2 dB (SPC)

100 kHz

freq.

100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB (SPC)

100 MHz < freq.

300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±2 dB (SPC)

1 M

input for Option 703, 704, 707, or 708

(using 50

feedthrough) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±3 dB

Note: Frequency response can be improved by calibration.