Supplemental characteristics
Supplemental characteristics are intended to
provide information useful in applying the instru-
ment by giving typical but non-warranted perform-
ance parameters. These are denoted as “typical”,
“nominal”, or “approximate”.
Key performance parameters for these test sets
are listed in the table. Standard (solid-state) and
Option 009 (mechanical) test sets have identical
performance except where noted below. Further,
these performance differences only affect the
measurement uncertainties of an “uncorrected”
8753C network analyzer system.
Parameter 85046A
85046B
85047A
Transmission Tracking
1
:
Magnitude
5
:
Standard:
±1.5 dB
±1.5 dB
±2 dB
Option 009:
±1.5 dB
±1.5 dB
±1.5 dB
Phase
6
:
±20 deg: 300 kHz to 2 MHz
±20 deg: 300 kHz to 2 MHz
±10 deg: 300 kHz to 3 GHz
±10 deg: 2 MHz to 3 GHz
±10 deg: 2 MHz to 2 GHz
±20 deg: 3 GHz to 6 GHz
Reflection Tracking
1
:
Magnitude
5
:
Standard:
±1.5 dB
±1.5 dB
±2 dB
Option 009:
±1.5 dB
±1.5 dB
±1.5 dB
Phase
6
:
±25 deg: 300 kHz to 2 MHz
±25 deg: 300 kHz to 2 MHz
±10 deg: 300 kHz to 3 GHz
±10 deg: 2 MHz to 3 GHz
±10 deg: 2 MHz to 2 GHz
±20 deg: 3 GHz to 6 GHz
Nominal Insertion Loss:
RF IN to Port 1, 2:
Standard:
14 dB +0.5 dB/GHz
19.5 dB +1 dB/GHz
4 dB +0.8 dB/GHz: 300 kHz to 3 GHz
17.5 dB +0.8 dB/GHz: 3 GHz to 6 GHz
Option 009:
12.5 dB +0.5 dB/GHz
18 dB +1 dB/GHz
2.5 dB +0.5 dB/GHz: 300 kHz to 3 GHz
16 dB +0.5 dB/GHz: 3 GHz to 6 GHz
RF IN to R:
18 dB +1.5 dB/GHz
18 dB +1.5 dB/GHz
19 dB +0.5 dB/GHz: 300 kHz to 3 GHz
34 dB +0.5 dB/GHz: 3 GHz to 6 GHz
Typical Isolation
100 dB
100 dB
100 dB: 300 kHz to 3 GHz
between Port 1 and 2:
90 dB: 3 GHz to 6 GHz
Nominal Impedance:
Port 1, 2:
50
Ω Nominal 75
Ω Nominal 50
Ω Nominal
RF IN, R, A, B:
50
Ω Nominal 50
Ω Nominal 50
Ω Nominal
DC bias input range:
+30 VDC, ±200 mA with some degradation in
+30 VDC, ±200 mA with no degradation
RF specifications (±500 mA maximum)
in RF specifications (±1 A maximum)
Connectors:
Port 1, 2:
7 mm
75
Ω Type-N(f)
7 mm
RF IN, R, A, B:
50
Ω Type-N(f) 50
Ω Type-N(f) 50
Ω Type-N(f)
DC bias input:
BNC(f)
BNC(f)
BNC(f)
5. Deviation from mean value.
6. Deviation from linear phase.