Test set specifications, 8753c system measurement uncertainty – Atec Agilent-85047A User Manual
Page 2
2
Achieve the highest system accuracy
with faster measurement speeds
The specifications listed in the table and the typi-
cal measurement uncertainties shown below apply
to all Agilent 8753C systems using either solid-
state switching (standard) or mechanical (Option
009) versions of the test sets.
Test set specifications
Specifications describe the instrument’s warranted
performance over the temperature range 0 to +55˚C
(except where noted).
8753C System measurement uncertainty
2
The graphs below show the typical measure-
ment uncertainty for the 8753C using full 2-port
error correction with the Agilent 85031B 7-mm
Calibration Kit.
Transmission measurements
3
Reflection measurements
4
Parameter 85046A
(50
Ω) 85046B
(75
Ω) 85047A
(50
Ω)
Frequency Range:
300 kHz to 3 GHz
300 kHz to 2 GHz
300 kHz to 3 GHz/3 MHz to 6 GHz
Directivity
1
≥35 dB: 300 kHz to 1.3 GHz
≥35 dB: 300 kHz to 1.3 GHz
≥35 dB: 300 kHz to 1.3 GHz
Port 1, 2:
≥30 dB: 1.3 GHz to 3 GHz
≥30 dB: 1.3 GHz to 2 GHz
≥30 dB: 1.3 GHz to 3 GHz
≥25 dB: 3 GHz to 6 GHz
Effective Source
≥14 dB: 300 kHz to 2 MHz
≥14 dB: 300 kHz to 2 MHz
≥20 dB: 300 kHz to 1.3 GHz
Match
1
:
≥20 dB: 2 MHz to 1.3 GHz
≥17 dB: 2 MHz to 1.3 GHz
≥16 dB: 1.3 GHz to 3 GHz
≥16 dB: 1.3 GHz to 3 GHz
≥16 dB: 1.3 GHz to 2 GHz
≥14 dB: 3 GHz to 6 GHz
Switch Repeatability:
±0.03 dB
1. Can be greatly improved by accuracy enhancement techniques.
2. These measurement uncertainty curves utilize an RSS model for the contribu-
tions of random errors, and a worst-case model for the contributions of dynamic
accuracy and residual systematic errors.
3. The graphs shown for transmission measurements assume a well-matched
device (S11=S22=0).
4. The graphs for reflection measurement uncertainty apply to either a one-port
device or a two-port device with more than 6 dB of insertion loss.