The specifications listed in the table and the typi-
cal measurement uncertainties shown below apply
to all Agilent 8753C systems using either solid-
state switching (standard) or mechanical (Option
009) versions of the test sets.

Test set specifications

Specifications describe the instrument’s warranted
performance over the temperature range 0 to +55˚C
(except where noted).

8753C System measurement uncertainty

2

The graphs below show the typical measure-
ment uncertainty for the 8753C using full 2-port
error correction with the Agilent 85031B 7-mm
Calibration Kit.

Transmission measurements

3

Reflection measurements

4

Parameter 85046A

(50

) 85046B

(75

) 85047A

(50

)

Frequency Range:

300 kHz to 3 GHz

300 kHz to 2 GHz

300 kHz to 3 GHz/3 MHz to 6 GHz

Directivity

1

≥35 dB: 300 kHz to 1.3 GHz

≥35 dB: 300 kHz to 1.3 GHz

≥35 dB: 300 kHz to 1.3 GHz

Port 1, 2:

≥30 dB: 1.3 GHz to 3 GHz

≥30 dB: 1.3 GHz to 2 GHz

≥30 dB: 1.3 GHz to 3 GHz

≥25 dB: 3 GHz to 6 GHz

Effective Source

≥14 dB: 300 kHz to 2 MHz

≥14 dB: 300 kHz to 2 MHz

≥20 dB: 300 kHz to 1.3 GHz

Match

1

:

≥20 dB: 2 MHz to 1.3 GHz

≥17 dB: 2 MHz to 1.3 GHz

≥16 dB: 1.3 GHz to 3 GHz

≥16 dB: 1.3 GHz to 3 GHz

≥16 dB: 1.3 GHz to 2 GHz

≥14 dB: 3 GHz to 6 GHz

Switch Repeatability:

±0.03 dB

1. Can be greatly improved by accuracy enhancement techniques.
2. These measurement uncertainty curves utilize an RSS model for the contribu-

tions of random errors, and a worst-case model for the contributions of dynamic
accuracy and residual systematic errors.

3. The graphs shown for transmission measurements assume a well-matched

device (S11=S22=0).

4. The graphs for reflection measurement uncertainty apply to either a one-port

device or a two-port device with more than 6 dB of insertion loss.