beautypg.com

Atec Aeroflex-6820A Series User Manual

Page 7

background image

For the ver y latest specifications visit

w w w.aeroflex.com

Bandwidth / Optional CF / DF (Q)
dB / Octave, dB / Decade Readout
Delta Marker On / Off
-1 dB gain compression

Fault Location

Find Max / Track Max
Next PK Right / Left
Set PK Level
Delta marker On / Off

General

Marker Table
Assign Active MKR / Position Active MKR
Set-up Markers (i.e. On / Off, Position)
Large Readout
All Off

Marker Resolution

Frequency: 6 digits or 1 Hz, user selectable
Power: 0.01 dB
Voltage: 1 nV

Measurement Manipulation

Display live measurement.
Display trace memory.
Display live measurement relative to trace memory.
Measurement hold may be applied for each trace.
Any input or ratio of inputs may be assigned to any one or more than
one trace(s). A trace may display absolute power, power relative to a
path calibration or power minus a trace memory.

Input Offsets

An offset in the range -99.99 dB to +99.99 dB in 0.01 dB steps may
be applied per detector input.

Weight - Variant and Option Dependent

16 kg (35 Ibs)

Size (Not including front handles)

230 mm H x 430 mm W x 570 mm D
9 in H x 17 in W x 22 in D

Power Supply

100-240 V~ (Limit 90-264 V~)
50-60 Hz (Limit 45-66 Hz)

108-118 V~ (Limit 90-132 V~)
50-400Hz (Limit 45-440 Hz)

200 W maximum

Rated Range of Use

Temperature

0 to 50°C

6825A and 6825AR only +5°C to 45°C only
Humidity

Up to 93% RH at 40°C

Conditions of Storage and Transportation

Temperature

-40 to +71°C

Humidity

Up to 93% RH at 40°C

Altitude

Up to 4570 m (15000 ft)

ELECTROMAGNETIC COMPATIBILITY

Conforms to the protection requirements of EEC Council directive
2004/108/EC.

Conforms to the limits specified in the following standards:

IEC/EN61326-1 : 2006

RF Emission Class A, Immunity table 3.

The radiated RF emission from this equipment is below Class A (refer-
ence CISPR 11). Class A equipment is intended for use in industrial
environments. There may be potential difficulties in ensuring electro-
magnetic compatibility in other environments due to radiated distur-
bances.

When using a 6230L detector, the noise floor may increase if exposed
to a conducted RF disturbance level of >1.5V.

SAFETY

Conforms with the requirements of EC Council Directive 2006/95/EC
(as amended) and the product safety standard IEC/EN 61010-1 :
2001 + C1 : 2002 + C2 : 2003 for class 1 portable equipment, for
use in a Pollution Degree 2 environment. The instrument is designed to
operate from an Installation Category 2 supply.