beautypg.com

Atec Aeroflex-6820A Series User Manual

Page 5

background image

For the ver y latest specifications visit

w w w.aeroflex.com

either trigger input or sync output depen-
dant upon trigger mode. TTL level

Options 025a & 025b Internal Pulse Modulator

Option 25a (6822A and 6823A)

Frequency Range

50 MHz to 18 GHz (8.4 GHz for 6822A)

Usable to 20GHz

RF Output Range

The levelled power range is reduced by:
<3.5 dB up to 6 GHz
<4.5 dB up to 14 GHz
<5.0 dB up to 18 GHz
when pulse modulation is selected

RF Level Accuracy Adds ± 0.3 dB to the levelled power

accuracy specification when pulse
modulation is enabled and for powers of
< -1 dBm

Source Harmonics (with Pulse Modulation enabled)
50 MHz - 2 GHz

<-35 dBc

2 GHz - 20 GHz

<-50 dBc

On/Off Ratio
50 MHz - 1 GHz

>55 dB

1 GHz - 9 GHz

>60 dB

9 GHz - 17 GHz

>70 dB

17 GHz - 18 GHz >80 dB
18GHz - 20GHz

>80 dB (typical)

Rise/Fall Times

(measured at 10% and 90%)

Rise Time

<8 ns (Typically < 5 ns)

Fall Time

<12 ns (Typically < 9 ns)

Option 25b (6824A, 6825A and 6825AR)

Frequency Range

50 MHz to 40 GHz (24 GHz for 6824A)

(46 GHz for 6825A)

RF Output Range

The levelled power range is reduced by:
<5 dB up to 20 GHz
<8 dB up to 30 GHz
<9 dB up to 40 GHz
when pulse modulation is selected

RF Level Accuracy Adds ± 0.3 dB to the levelled power accuracy
specification when pulse modulation is enabled and for output powers
of <-3 dBm

Source Harmonics (with Pulse Modulation enabled)
50 MHz - 375 MHz

<-35 dBc

375 MHz - 24 GHz

<-50 dBc

24 GHz - 40 GHz

<-20 dBc

On/Off Ratio
50 MHz - 10 GHz

>60 dB

10 GHz - 26.5 GHz

>60 dB (typically > 70 dB)

26.5 GHz - 40 GHz

>60 dB (typically > 80 dB)

Rise/Fall Times

(measured at 10% and 90% of edge)

Rise Time

<7 ns (Typically < 6 ns)

Fall Time

<11 ns (Typically < 10 ns)

Pulse Modulation Control

Modes
Pulse, Pulse CW
External (via rear panel BNC connector)
Internal (if Opt 23 fitted)

Control

Control of pulse modulation is:

Internal via soft key menu when the modulation generator
option (Opt 023) is fitted or

External via the rear panel BNC Mod in/out socket.

Level is TTL, High = On, Low = Off.

When pulse mod Off is selected the output is the selected CW
output level

Pulse CW In both internal or external modes, allows setting of

output Level in the 'On' condition for reference or
calibration.

SCALAR ANALYZER

SYSTEM FEATURES

Frequency Range

As per source frequency range

Number of Inputs

3 detector inputs

Number of Measurement Points

User selectable from 2 to 1601

Applications

Return loss vs frequency
Insertion loss vs frequency
Fault Location
Voltage vs frequency

Detection Modes

AC and DC

Noise Reduction

Averaging, 1 to 1000
Smoothing, 0.01 to 20%

Power Measurements

Using scalar detectors

Detector Correction

Frequency response and linearity read from EEPROM for 6230A/L and
fault locators.
Support for 6230 and autotesters.

INSERTION LOSS MEASUREMENTS

Measurement Dynamic Range, AC Scalar Detection, with 623XA
Detector

Max source output to -60 dBm
Max source output to -65 dBm (with averaging)

Typical values:

>65 dB (10 MHz to 40 GHz)
>75 dB (1 MHz to 3 GHz) only with 6232A

Measurement Update Rate

401 points in 270 ms with DC detection

Calibration

Through path calibration or short and short/open calibrations for single
ended insertion loss

Inputs

Single input or ratio

Accuracy

Linearity + mismatch

Linearity (applies after normalization)

Linearity (for Power Levels >-50 dBm)

±0.2 dB / 10 dB but not >0.5 dB in total