General information
Temperature range
Operating: 0 to 55 °C
Storage: –40 to 75 °C
Power requirements
48 to 66 Hz, 100/120/220/240 V ±10%,
typically 155 VA
Dimensions: 178 H x 425 W x 482 mm D
(7.0 x 16.75 x 19.0 in)
Weight: Net 22 kg (48 lb), Shipping: 28 kg (61.5 lb)
Power calibrator (Option 002 only)
The 8757D’s internal power calibrator provides
a 50 MHz reference standard for characterizing
the absolute power accuracy and dynamic power
accuracy of the 85037 series precision detectors.
Frequency: 50 MHz ±0.2 MHz
Output power: (25 ±5 °C)
Range: +20 to –50 dBm
Accuracy at 0 dBm: ±0.05 dB
Linearity: (over any 10 dB range)
±0.08 dB (+20 to +10 dBm)
±0.04 dB (+10 to –30 dBm)
±0.06 dB (–30 to –50 dBm)
SWR: ≤1.05
Modes of operation
DC Mode (unmodulated)
AC Mode (modulated at 27,778 Hz ±12 Hz)
Connector: Type-N (f)
Accessory included
A Type-N (m) to 3.5-mm (f) adapter is provided
to allow calibration of the 85037B (3.5 mm)
precision detector.
Ordering information
Agilent 8757D scalar network analyzer (select one)
8757D-700 Standard
network
analyzer
8757D-001
Adds fourth detector input
8757D-002
Adds internal power calibrator
8757D-012
Adds fourth detector and internal
calibrator