DATA SHEET > CWS 500N2 > 20120604

TECHNICAL DETAILS

GENERAL DATA

Dimensions,
weight

19"/6HU, approx. 31kg

Supply voltage

115V or 230V +10/-15%, 50/60Hz

Input power

Max. 380W

Power factor

cos(phi) = 0.96 at max. output power
as per IEC 555

Fuses

2x6.3AT (115V) or 2x3.15AT (230V)

Cooling

Active cooling, air ventilation

Temperature

10°C - 40°C

Rel. humidity

Max. 85%, non-condensing

ACCESSORIES

Cables

N-type coaxial cables to connect the
3dB attenuator and/or the injection
probe, BNC-coaxial cable to connect
the current monitoring probe, with
N-to-BNC adapter

icd.control

Extensive and most versatile remote
control and reporting software.
The standard library helps to
configure the test setup.
Multiple interruption functions
automated by IEEE instruments or
manually.
Easy-to-use as well as expandable to
complex test routines based on
vector definitions.

OPTIONS

ATT3/100

3dB attenuator, 100W

ATT20/15

20dB attenuator for current monitor
path, the set includes 2 units

ATT20/100

20dB attenuator, 100W
for small level RF signals as per
MIL STD 461 an DO-160

T-50A

50ohm, 6W termination resistor

Calibration

Adaptors and cal jigs

CDNs

as per IEC 61000-4-6 (refer to
seperate list)

Clamps

EM clamp as per IEC 61000-4-6
Current injection clamps
Current monitoring clamps

R-100x

150ohm-to-50ohm matching
impedance

www.emtest.com