DATA SHEET > CWS 500N2 > 20120604
TECHNICAL DETAILS
BULK CURRENT INJECTION AS PER ISO 11452-4
Output level
As required in ISO 11452-4, using
closed loop or substitution method
BULK CURRENT INJECTION AS PER MIL 461 CS 114
Output level
As required in MIL 461 CS 114, using
the closed loop method
MEASUREMENTS FOR BULK CURRENT INJECTION
Directional
coupler
Included to measure forward power
and reverse power
Forward power
Internal power meter #1
Reverse power
Internal power meter #2
Injected current
Internal power meter #3
TEST ROUTINES FOR BULK CURRENT INJECTION
ISO 11452-4
Operation via icd.control
MIL 461 CS114
Operation via icd.control
IEC 61000-4-6
Output level
1V - 30Vrms (emf)
all standard test levels are
guaranteed with all coupling
methods
MEASUREMENTS, IEC 61000-4-6
Cal in (BNC)
Integrated power meter to record the
calibration data of a coupling device
Injected current
Measured by internal power meter
RF indicator
LED indicating the RF output status
LCD
Display of the test level and the
preselected frequency value
TEST ROUTINES FOR IEC 61000-4-6
Quick Start
Immediate start; easy-to-use and
fast test routine
User Test routines
Frequency sweep
Service
Service, Set-up
OUTPUT
RF output
N connector at the front panel
Ouptut power
Built-in amplifier
50dBm (9kHz - 400MHz) at 1dB
compression.
Gain amplifier
> 50dB
Output impedance
50ohm
Harmonic
distortion
< -20dBc at max. power
Insertion loss
Approx. 1dB (directional coupler +
RF relay)
TIME PARAMETERS
Dwell time
td = 0.3s - 9,999s
Pause time
tr = 0/0.3s - 9,999s
TEST FREQUENCIES
Frequency range
9kHz - 400MHz (built-in amplifier)
9kHz - 1,000MHz (ext. amplifier)
Unmodulated
signal
CW (continuous wave)
Amplitude
modulation
1kHz, 80% AM (EN/IEC 61000-4-6)
1kHz, 95% AM (automotive)
400Hz, 80% AM
50Hz, 80% AM (automotive)
2Hz, 80% AM (IEC 60601-1-2)
Pulse modulation
1Hz, 50% duty cycle (EN 50130-4)
1kHz, 50% duty cycle (MIL 461)
INTERFACE
Serial interface
USB
Parallel interface
IEEE 488, addresses 1 - 30
Fail 1
BNC input; test will be stopped
(active low)
Fail 2
BNC input; test status will be saved
(max. 10 events) when active low.
Test continues