DATA SHEET > CWS 500N1 > 20120604

TECHNICAL DETAILS

TEST LEVEL

Output level

1V - max. 30V (emf)
all standard test levels are
guaranteed with all coupling
methods

Output power

80W (nominal)

Output impedance

50ohm

Harmonic
distortion

< -20dBc at max.power

TEST FREQUENCIES

Sinus (CW)

100kHz to 300MHz

Frequency bands

100kHz to 9.999MHz
10MHz to 99.99MHz
100MHz to 300MHz
in the Quick Start menu the step size
can be selected by the operator

Unmodulated
signal

CW (continuous wave)

Amplitude
modulation

1kHz, 80%AM as per IEC 61000-4-6
2Hz, 80%AM as per IEC 60601-1-2
400Hz, 80%AM

Pulse modulation

1Hz, 50% duty cycle as per
EN 50130-4

MEASUREMENTS

Monitor

Integrated RF power meter,
measuring input for CDN and clamp
calibration as well as current
monitor for clamp applications

RF indicator

LED indicating the RF output status

LCD

Online display of the test level and
the preselected frequency value

Cal data F1 - F5

5 internal memories to save
calibration data,
1% step width as required by the
standard

TIME PARAMENTERS

Dwell time for CW
& AM

td = 0.3s to 9,999s

Dwell time for PM

td = 3s to 9,999s

Pause time

tr = 0/0.3s to 9,999s

OUTPUT

Direct RF output

BNC (on the front panel)

TEST ROUTINES

Quick Start

Immediate test start; easy-to-use
and fast

User Test routines

Voltage sweep
Frequency sweep
Dwell time sweep

Standard Test
routines

Level 1 to Level 3 (IEC 61000-4-6)
Automatic Level X to Level Y

Cal procedure

Calibration of the complete test
set-up, calibration data saved in
internal memory

Service

Service, set-up

SIGNAL GENERATOR

Output level

-55dBm to 0dBm

Frequency range

9kHz to 1GHz

Output impedance

50ohm

Direct RF output

To control an external amplifier

MEASUREMENTS FOR BCI APPLICATION

PM 1000

3-channel power meter up to 1GHz
- to measure the Forward power
- to measure the Reverse power
- to measure the injected current

INTERFACE

Serial interface

USB

Parallel interface

IEEE 488, addresses 1 to 30

Fail 1

BNC input; test will be stopped when
active low

Fail 2

BNC input; test status will be saved
(max. 10 events) when active low.
Test will continue.

GENERAL DATA

Dimensions,
weight

19"/3HU, approx. 17kg

Supply voltage

115V to 230V +10/-15%, 50/60Hz

Input power

Max. 380W

Power factor

cos (phi) = 0.98 at max. output
power as per IEC 555

Fuses

2 x 6.3AT (115V) or 2 x 3.15AT
(230V)

Cooling

Active cooling, air ventilation

Temperature

10°C to 40°C

Rel. humidity

Max. 85%, non condensing

www.emtest.com