Options, Specifications, Environmental specifications – Atec Agilent-U9391C-F-G User Manual
Page 5: Output connectors
5
Option FFF – female (output port)
Option FFM – male (output port)
Specifications refer to the performance standards or limits against which the
U9391C/F/G comb generators are tested.
Typical characteristics are included for additional information only and they are not specifi cations.
These are denoted as “typical”, “nominal” or “approximate” and are printed in italic.
Options
Output connectors
Specifications
1
Environmental
Specifications
Specifi cations
U9391C
U9391F
U9391G
Output frequency range
10 MHz to 26.5 GHz
10 MHz to 50 GHz
10 MHz to 67 GHz
Input frequency range
10 MHz to 6 GHz
10 MHz to 6 GHz
10 MHz to 6 GHz
Input power range
-15 to +15 dBm
-15 to +15 dBm
-15 to +15 dBm
Min output power per picket
at 10 MHz Input PRF
-80 dBm
-95 dBm
-100 dBm
Amplitude fl atness vs. output frequency
at 10 MHz Input PRF
< 12 dB
< 25 dB
< 40 dB
Amplitude fl atness vs. input power
0.1dB (typical)
0.5dB (typical)
1.0 dB (typical)
Phase fl atness/degree
10 MHz to 3 GHz
3 GHz to 20 GHz
20 GHz to 26.5 GHz
26.5 GHz to 28 GHz
28 GHz to 38 GHz
38 GHz to 45 GHz
45 GHz to 50 GHz
50 GHz to 67 GHz
± 8.5
± 6.5
± 8.5
+ 10 /- 10
+ 10 /- 10
+ 10 /- 10
+ 10 /- 10
+20 / -10
+20 / -15
+20 / -15
+ 10 /- 10
+ 10 /- 10
+ 10 /- 10
+ 20 /- 10
+20 / -10
+20 / -15
+20 / -15
+35 / - 35
Pulse width
< 23 ps
< 23 ps
< 23 ps
Divide ratio
1,2,4,8,16
1,2,4,8,16
1,2,4,8,16
Input return loss, S11
10 MHz to 6 Ghz
> 10 dB
> 10 dB
> 10 dB
Output return loss, S22
10 MHz to 10 GHz
10 GHz to 20 GHz
20 GHz to 26.5 GHz
26.5 GHz to 45 GHz
45 GHz to 50 GHz
50 GHz to 67 GHz
> 10 dB
> 10 dB
> 10 dB
> 10 dB
> 10 dB
> 7 dB
> 7 dB
> 5 dB
> 10 dB
> 10 dB
> 7 dB
> 7 dB
> 5 dB
> 5 dB
Specifi cations
Limits
Temperature
Operating
Storage
Error corrected range
Cycling
0 to +40 °C
–40 to +70 °C
23 °C ±3 °C
–65 to +85 °C, 10 cycles at 20 °C per minute. 20 minutes dwell time per MIL-STD-883F, Method 1010.8,
Condition C (modifi ed)
Relative humidity
Operation
Storage
50% to 95% RH at 40 °C, 24 hours cycling, repeated 5 times
90% RH at 65 °C, one 24 hour cycle
Shock
End-use handling shock
Transportation
shock
Half-sine waveform, 2-3 ms duration, 60 in/s (1.6 ms) delta-V
Trapezoidal waveform, 18-22 ms duration, 337 in/s (8.56 ms) delta-V
Vibration
Operating
Survival
Random: 5 to 500 Hz, 0.21 grms, 10 min/axis
Random: 5 to 500 Hz, 2.09 grms, 10 min/axis
Swept sine: 5 to 500 Hz, 0.5 grms, 10 min/axis, 4 resonance search, 10 min dwell
Altitude
Operating
Storage
< 4,572 meters (15,000 ft)
< 15,000 meters (50,000 ft)
ESD immunity
Direct
discharge
1
Air
discharge
8.0 kV per IEC 61000-4-2
15 kV per IEC 61000-4-2
1. To outer conductor
U9391C/F/G comb generators are designed to fully comply with Agilent’s product
operating environment specifications. The following are the summarized
environmental specifications for these products.
1. Note: The U9391C/F/G was specially designed for use with the PNA-X ONLY.