Technical data
Power supply
85 V … 264 V (47 Hz … 63 Hz)
Power consumption
max. 60 VA
Internal rechargeable battery
2 x 6 V / 6.5 Ah - approx. 8 h self-supporting operation
Output voltage
0 … 75 kV
eff
symmetrical
For insulating liquids with tan
δ
values
< 4.5 or specific resistance
ρ
> 30 MΩm
Voltage slew rate
0.5…10 kV/s
Switch-off time
< 10 µs
Voltage control
Real Breakdown Monitoring
Internal measurement of the oil sample temperature
0 … 99 °C
Temperature resolution
1 °C
Display
LCD colour display (320x240)
Accuracy
0 … 75 kV ± 1 kV
Resolution
0.1 kV
Languages
German, English, French, Spanish, Italian, Portuguese (Br), Portuguese
(Pt) Dutch, Polish, Russian, Chinese (Cn), Chinese (Tw), Czech
Pre-programmed standards
ASTM D 1816-04 1 mm, ASTM D 1816-04 2 mm, ASTM D 877 / 87 PA,
ASTM D 877 / 87 PB, BS EN 60156, CEI EN 60156, CSSR RVHP / 85,
IEC 60156 / 1995-05, IRAM 2341 / 72, JIS C2101 / 78, PN 77 / E-04408,
PA SEV EN 60156, UNE EN 60156, NF EN 60156, SABS EN 60156,
VDE 0370 Part 5 / 96, AS1767.2.1, Individual measurement
Customer-specific programs
10
Interface
USB 2.0
Printer
Matrix printer, 24 characters, 57 mm Normal paper
Ambient temperature
-10 … +55 °C
Storage temperature
-20 .. +60 °C
Relative humidity
Non condensing
Dimensions (W x H x D)
476 x 372 x 340 mm (closed)
476 x 635 x 420 mm (open)
Weight without options
29 kg
Degree of protection
IP 32
Designed and manufactured in acc. with the following
standards
IEC 61010-1, EN 61010-1, VDE 0411 Part 1
Delivery includes
Test vessels
Options
Oil tester DPA incl. Printer and
battery
One test vessel
Magnetic oil stirrer
Lifting stick for magnetic stirrer
A setting gauge
Mains connection cable
User manual
Shoulder strap
Test vessel 0.4 l in acc. with
IEC 60 156 Fig. I with cover
Test vessel 0.4 l in acc. with
IEC 60 156 Fig. II with cover
Test vessel 0.4 l in acc. with
ASTM D 877 with cover
Test vessel 0.4 l in acc. with
ASTM D 1816-04 with cover
PC interface and BAUR Oil testing
software ITS 1.0 incl. Cable
Robust, completely sealed cover
bag (operation in bag possible)
Setting gauges
1/2/2.5/2.54/4/5 mm
Ring nut wrench for disassembling
the test vessel
Data sheet