Baur oil tester dpa 75 c, Features
BAUR Oil tester
DPA 75 C
The portable BAUR oil tester DPA automatically measures the electrical breakdown strength of
insulating liquids fully.
Features
Compact and robust device for
portable use onsite
Test voltages from 0 to 75 kV
eff
Repeatable breakdown
measurements in mineral, silicone
and plant oils
Reliable measurement results due
to shortest switch-off time
(<10 µs)
Self-sufficient operation possible
due to integrated battery or mains
operation
User interface in 13 languages
Functional design for high
efficiency, user-friendliness and
safety during full mobility
Clear breakdown detection with
RBM technology and a new
measurement principle directly in
the HV part
Automatic self-test with HV output
voltage test
Fully automatic measuring
sequences from 18
pre-programmed test standards
and 10 user-programmable test
sequences
Built-in measurement of the
temperature of the insulating liquid
Ergonomic operating unit with
oil-proof membrane keyboard, easy
to read LCD colour display and
integrated printer
EMC shield for prevention of
damage to EDP systems
Measurements without air inlet due
to high quality test vessels
(compatible with DPA 75 vessels)
Precise setting of the electrode gap
Transport protection for test vessel
PC interface including software for
controlling, recording and archiving
measurements (option)
Data sheet