Display, Operation, Mechanical design – Atec Dranetz-Profitest User Manual
Page 3: Interfaces, Pc software features
GMC-I Messtechnik GmbH
3
PROFITEST PV
Peak Power Measuring Instrument and Curve Tracer for
PV Modules and Strings (Measurement at Capacitive Load)
Display
Display
Color LCD with LED backlight
Resolution
480 x 272 pixels
Suitable for use in sunlight
Operation
•
Menu driven via touch-screen directly at the instrument
•
Operation and evaluation alternatively with Windows software
•
Connection to PC: USB, standard B socket
•
USB cable: standard USB 2.0 cable
Mechanical Design
Protection
IP20
Dimensions
W x H x D: 480 mm x 315 mm x226 mm
Weight
Approx. 9,5 kg
Interfaces
The PROFITEST PV is equipped with the following inputs and out-
puts (except for the jacks for the external power pack, all inter-
faces are located on the front panel, and are labeled):
PC Software Features
PV Analyzer
Software for visualization, evaluation and documentation of mea-
sured characteristic curve values with database
•
Measured characteristic curve values are read in from the
PROFITEST PV
•
Graphic representation of the characteristic I-U curve
–
With calculated MPP – maximum power point (Pmax)
–
In comparison with the characteristic power curve
–
In comparison with the TRMS curve
–
In comparison with the STC curve
–
Characteristic I-U curve with display of measuring points
•
Representation of measured and calculated values under STC
•
Overview of characteristic I-U curves for a given test series in
browser window
•
Export of measured values or results (e.g. XLS file)
•
Generation of test reports (e.g. PDF file)
•
Online measurement – graphic representation of the charac-
teristic curve and measured values (also suitable for continu-
ous measurement)
•
Online access to the database and file management at the
PROFITEST PV
•
Compatible with MS Windows
®
NT, 2000, XP, Vista, 7
Interface
Function
Mains
External power pack: connection via jack plugs,
5.5 x 2.1 mm
Temperature
Connection for temperature sensor
• Instruments with 2 temperature measuring inputs:
external Pt100 for acquiring temperature at the back
of the module
• Other instruments: external Pt100 or Pt1000
(depending on variant) for measuring temperature at
the back of the reference cell
Irradiation
Connection for irradiation reference sensor (Phox)
• The Pt100/Pt1000 reference sensor and the mea-
sured irradiation value are combined into a single 8-
pin plug.
4-wire measurement
Measurement input (voltage measurement)
Current input
Power input (for current measurement)
PC
Connection via USB cable