Terrasas - specifications – Atec Elgar_TerraSAS User Manual
Page 5
![background image](/manuals/318104/5/background.png)
858.458.0223
Specifications
AC Power
AC Input Voltage: 115V (for DCS) 208VAC three phase Std, 400VAC and 480VAC three phase are optional
(Input current depends on power rating)
DC Output
Open circuit voltage, Voc: 0 - 1000VDC
Short circuit current, Isc: 0 – 1000A
Maximum output power at MPP: 1MW
(Lower voltage ranges will provide proportionately higher currents) 1-6 channel output, consult factory for additional channels.
Programmable Parameters
Irradiance level: 0 to 2000 W/m²
Temperature: -40 to 90°C
Temperature Coefficient: 0 to -65,000 mV/°C
Simulation Times: 0 to 65,000 seconds
Isolation relay and polarity relay closure
Accuracy
Voltage Readback: 0.2% of rated max voltage
Current Readback: 0.5% of max current
Programming Interface
Ethernet with RJ-45 connector / LAN
AC Input Connections
Finger safe, pressure type connectors three phase AC four wire plus safety ground stud AC input circuit breaker
DC Output Connections
Finger safe, pressure type connectors positive and negative
Safety
The output isolation relay operates as a disconnect relay in the event of a malfunction or an open interlock contact
Output Voltage and Current Ranges
Power (MPP)
1 kW
5 kW
10 kW
15 kW
RMS
P-P
DC Leakage Current
Voc
80V
Isc=12.5
Isc=83A
Isc=167A
Isc=250A
20 mV
100 mV
600Voc
Isc=1.6A
Isc=8A
Isc=16A
Isc=25A
60 mV
350 mV
335mA
1000Voc
N/A
Isc=5A
Isc=10A
N/A
<1mA
MMPT
Scalable (MPPT)
1000W to 1.0MW
Response to MPPT
Up to 120Hz
Current Slew Rate
3msec/A
Control Loop Sampling Rate
1usec / 10kHz
TerraSAS - Specifications
Programmable Setpoints
Voc
0-Rated output voltage
Fill Factor
0.5 to 0.95
Vmp
0-Voc
Imp
0-Isc
ISC
0-Rated output current
Over Voltage Protection (OVP)
0.1% to 110% of Voc Max
Static and Dynamic Programmable PV Array Parameters
Irradiance Level
0-2,000W/m2
Temperature
-100 to +100*C
Voltage Level
0-600/1,000V 80V - Consult factory for other voltages
Current level to rated output current
0-Rated Output (see MPP Chart)
Voltage Temperature Coefficient
0 to -2% / *C
Arbitrary VI Curve
Up to 4096 data points
VI Curve Set Point Accuracy
Voltage
<0.1%, FS
Current
<0.5%, FS
Programming Resolution
Programming Resolution
<0.002% of FS
Voltage / Current
<0.002% of FS
5