Specifications, Dielectric – Atec AEMC-CA6121 User Manual
Page 4
■
C.A 6121
(Great Britain model)
................P01.1456.02
■
C.A 6121
(Continental Europe model)
........P01.1456.01
Instruments supplied with :
1 mains lead, 3-core,
1 set of 2 keys to lock the instrument,
2 high voltage guns (2 m cable),
1 discharge cable,
2 continuity leads 2.5 m, with double
contact crocodile clips,
2 safety leads 3 m (red/ black),
2 crocodile clips (red/ black),
1 test prod (red), 1 carrying bag
for the accessories and the user manual
Additional accessories :
■
PC software
« C.A 6121 TRANSFER » ....P01.1019.15
supplied with cable RS232 DB9M/DB25F
+ adapter DB25M/ DB9F
■
Serial printer ....................P01.1029.01
supplied with cable RS232 DB9M/ DB9M
■
Remote control pedal ......P01.1019.16
■
Signalling lights ................P01.1019.17
■
2 high voltage guns
(fitted with cable 6 m) ........P01.1019.18
Specifications
DIELECTRIC
Nominal test voltage (50 Hz)*
1000 V
1250 V
1500 V
Triggering current
1, 3, 5, 10, 20, 30, 40, 50, 60, 70,
80, 90, 100, 150, 200, 250, 300,
350, 400, 450, 500 mA
1, 3, 5, 10, 20, 30, 40, 50, 60,
70, 80, 90, 100, 150, 200, 250,
300, 350, 400 mA
1, 3, 5, 10, 20, 30, 40,
50, 60, 70, 80, 90, 100, 150,
200, 250, 300, 333 mA
* Test voltage greater than 1000, 1250 or 1500 V / 50 Hz at a voltage U
mains
= 230 V and a power P
load
= 500 VA
Display
Test voltage
Test current (sinusoidal)
Range
0...2.00 kV
0...199.9 mA
200...500 mA
Resolution
10 V
0.1 mA
1 mA
Accuracy
±(2% L + 20 V)
±(2% L + 0.3 mA) *
* ± (2% rdg + 0.5 mA) if triggering current set to 30, 40, 50, 60, 70, 80, 90 or 100 mA and ± (2% rdg + 2 mA) on 150, 200, 250, 300, 330, 350, 400, 450 or 500 mA
■
Display of the character of the test current : resistive, capacitive or absolute value
■
Timer : without or adjustable from 1 s to 5 min
INSULATION RESISTANCE
CONTINUITY
DISCHARGE TIME
GENERAL SPECIFICATIONS
Display
Insulation resistance
Test voltage
Range *
0...1.999 M
Ω
2.00...19.99 M
Ω
20.0...199.9 M
Ω
200...500 M
Ω
0...1200 V
Resolution
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
1 V
Accuracy
±(2% L + 2 pt)
±10% L
±(2% L + 2 pt)
* automatic ranges
Display
Voltage drop (
∆
U)
Resistance (R
10A
) *
Test voltage
Test current
Range
0...10 V
0...999 m
Ω
1.00...1.99
Ω
0...12 V
0...9.99 A
10.0...25.0 A
Resolution
0.01 V
1 m
Ω
10 m
Ω
0.01 V
0.01 A
0.1 A
Accuracy
±(2% L + 0.02 V)
±(2% L + 2 m
Ω
)
± 5% L
±(2% L + 0.02 V)
±(5% L + 2 pt)
±3% L
* automatic ranges
■
Rated voltage : 500 or 1000 V ...
—
■
Measurement current : 1 mA min
■
Measurement with or without adjustable threshold value (0.2...500 M
Ω
)
■
Timer : without or adjustable from 2 s to 10 min
Display
Discharge time
Range
0...10 s
Resolution
0.1 s
Accuracy
±(2% L + 0.2 s)
■
Level of safety voltage : 60 V
■
Automatic threshold of discharge time :
1 s at the power supply terminals (2 point measurement)
5 s at the internal components (4 point measurement)
■
Max service voltage: 600 V AC/DC
■
Mains voltage : 230 V / 50 Hz
■
Max power absorbed : 600 VA
■
LCD screen : dot matrix 128 x 64 with
permanent backlighting and adjustable contrast
■
Interface : RS232
■
Memories : 999 values
■
Remote control signals : On / Off and Record
■
Working temperature : 0...50°C
■
Working humidity : 85% RH max (0...+40°C)
■
Dimensions (W x D x H) : 400 x 260 x 250 mm
■
Weight : 11 kg
■
Protection index : IP 40 (open) IP 54 (closed)
■
Protection of measurement circuits : by HBC
fuses
■
Electrical safety : IEC 1010-1 Cat III 600 V
Applicable standards :
EN 60204-1 “ Test of electric machines ”
VDE 104 “ Installation and operation of electric
test equipment ”
EN 61180-1 and -2 “ High voltage test techniques
for low voltage equipment ”
To Order
■
Voltage drops (
∆
U) max pre-set, depending on the cross section of the cable :
Cross-section of the cable tested
0,5 mm
2
0,75 mm
2
1 mm
2
1,5 mm
2
2,5 mm
2
4 mm
2
6 mm
2
Max voltage drop
5 V
5 V
3.3 V
2.6 V
1.9 V
1.4 V
1.0 V
■
Measurement of low resistance (R
10A
) with or without adjustable
threshold value (10...2000 m
Ω
)
■
Max output voltage : 12 V~
■
Timer : without or adjustable from 1 to 15 s
■
Measurement
current (0...0.5
Ω
) : > 10 A ~
■
Connection system : 4-wire
FRANCE
190, rue Championnet
75876 PARIS Cedex 18
Tél. : (33) 01 44 85 44 85
Télex 269816 - Fax : (33) 01 46 27 73 89
7
S U B S I D I A R I E S
W O R L D - W I D E
906 111 632 - Ed.1 - 11/97 Non contr
actual document - Confir
m specifications bef
ore order
ing - Photos Blaise PAST
OR
-
- 01 41 41 08 94
UNITED KINGDOM
Waldeck House - Waldeck Road
Maidenhead SL6 8BR
Tel. : 1628 788 888
Fax : 1628 628 099