beautypg.com

Specifications, Ordering information – Atec Chroma-19032 User Manual

Page 8

background image

19032

: Electrical Safety Analyzer

19032-P

: Electrical Safety Analyzer (500VA)

A190301

: 8HV Scanning Box

A190302

: 5HV/3GC Scanner

A190303

: 3HV/5GC Scanner

A190304

: 8HV Scanner

A190305

: Line Leakage Current Scanner (generally)

A190306

: Hipot/Line Leakage/Probe Scanner (10A)

A190307

: L-N Scanner & Leakage Current Scanner

A190308

: Hipot/Line Leakage/Probe Scanner (20A)

A190313

: 500VA Isolation Transformer

A190314

: 1000VA Isolation Transformer

A190316

: Dummy Load

A190317

: BarCode Scanner

A190508

: GPIB Interface for 19302

A190334

: Ground Bond 40A for 19032

A190337

: Ground Bond 60A for 19032

A190338

: 19001 EST Software

A190343

: 19" Rack Mount Kit for 19032

A190344

: HV Gun

A190349

: Universal Corded Product Adapter

A190350

: Hipot/Line Leakage/

Probe Scanner (20A) 6000-08

A190353

: 4HV/4GC Scanner

A190355

: 19" Rack Mount Kit for 19032-P

A190356

: GPIB Interface for 19302-P

A190708

: ARC Verification Fixture

SPECIFICATIONS

Distributed by:

Worldwide Distribution and Service Network

19032-E-201102-2000

CHINA

CHROMA ELECTRONICS

(SHENZHEN) CO., LTD.

8F, No.4, Nanyou Tian An

Industrial Estate, Shenzhen,

China PC: 518054

Tel: +86-755-2664-4598

Fax: +86-755-2641-9620

U.S.A.

CHROMA ATE INC. (U.S.A.)

7 Chrysler Irvine, CA 92618

Tel: +1-949-421-0355

Fax: +1-949-421-0353

Toll Free: +1-800-478-2026

CHROMA ATE INC.

致茂電子股份有限公司

HEADQUARTERS

66, Hwa-Ya 1st Rd., Hwa-Ya

Technology Park, Kuei-Shan Hsiang,

Taoyuan Hsien 33383, Taiwan

Tel: +886-3-327-9999

Fax: +886-3-327-8898

http://www.chromaate.com

E-mail: [email protected]

Developed and Manufactured by :

EUROPE

CHROMA ATE EUROPE B.V.

Morsestraat 32, 6716 AH Ede,

The Netherlands

Tel: +31-318-648282

Fax: +31-318-648288

ORDERING INFORMATION

*Different option have different specification.

Model

19032

19032-P

Mode

AC/DC/IR/GB/LC

Withstanding Voltage Test

Output Voltage

DC:0.05 ~ 6kV , AC : 0.05 ~ 5kV

Load Regulation

±(1% reading +0.1% range) ±(2% reading +0.1% range)

Voltage Resolution

2V

Voltage Accuracy

±(1% reading +0.1% range) ±(2% reading +0.1% range)

Cutoff Current

DC : 20mA , AC : 40mA

DC : 20mA , AC : 100mA

Current Resolution

0.1 μA DC ; 1μA AC

Current Accuracy

±(1% reading +0.1% range) ±(2% reading +0.5% range)

Output Frequency

50Hz / 60Hz

Test Time

0.3 ~ 999 sec , continue

Ramp Time

0.1 ~ 99.9 sec, Off

Fall Time

0.1 ~ 99.9 sec, Off

Waveform

Sine wave

Insulation Resistance Test

Output Voltage

DC : 0.05 ~ 1kV

Voltage Resolution

2V

Voltage Accuracy

±(2% of reading +0.5% of range)

IR Range

1MΩ ~ 50GΩ

Resistance Resolution

0.1MΩ

Resistance Accuracy

5% typical

Ground Bond Test

Output Current

AC : 1 ~ 30A

AC : 3 ~ 40A

Current Accuracy

±(1% of reading +0.2% of range)

GR Range

10mΩ ~ 510mΩ

Resistance Resolution

0.1mΩ

Resistance Accuracy

± (1% of reading + 0.1% of full scale)

Test Method

4 wires

Flashover Detection

Setting Mode

Programmable setting

Detection Current

AC : 20mA, DC : 10mA

Secure Protection Function

Ground Fault Interrupt

-

0.5mA ±0.25mA AC

Floating Output

-

3.5mA, front output only

Panel Operation Lock

Present password

Interlock

YES

GO/NG Judgment Window

Indication,Alarm

GO : Short sound,Green LED ; NG : Long sound, Red LED

Data Hold

Least tests data memories

Memory Storage

50 setups with up to 100 groups recall

Interface

9pin D-sub I/O control / RS232 / GPIB (Optional) / USB TMC (19032-P)

General

Operation Environment Temperature : 0°C ~ 40°C, Humidity : ± 80 % RH

Power Consumption

No load : < 100 W With Rated

load : 800 W

No load : < 100W

Rated load : 1000W

Maximum load : 1200W

Power Requirements

90~132Vac or 198~264Vac, 47~66Hz

Weight

Approx. 20KG(19032) 25KG (19032-P)

Option

6000-04 ~ 08*

Support Mode

AC/DC/IR/LC

DUT Input Power

Capacity

AC : 300V / 10A / 20A max.

Short Protection

20A, 250V fuse for DUT shorted.

Measurement Mode

Input Characteristic

DC - 1MHz

Input Impedance : 1M//20pF

Measurement Mode

Normal, Reverse,

Single Fault Normal,

Single Fault Reverse

Measurement

Devices

(Five measure

device)

UL 544 NP ; UL 544 P

UL 1563 ; UL 60601-1,

IEC60601-1;

UL 3101-1, UL/IEC 60950,

UL 1950-U1* ; UL 2601-U1*

IEC60990

Probe Connection

Line to Ground,

Line to P2,

P1 to P2

HI-LO Limit
LC HI-LO Limit

0 ~ 9.99mA, 1μA resolution

Current HI-LO Limit

0 - 19.99Amp*

VA HI-LO Limit

0 - 4400VA

VA Resolution

0.1VA

JAPAN

CHROMA JAPAN CORP.

472, Kohoku-ku Yokohama

city Ntsupa 223-0057 Japan

Tel: +81-045-542-1118

Fax: +81-045-542-1080