Ac/dc/ir, Hipot, Tester – Atec Chroma 19071 User Manual
Page 2: Chroma ate inc, Specification application order information, Panel description
![background image](/manuals/318085/2/background.png)
7
8
11
10
9
12
All specifications are subject to change without notice.
Developed and Manufactured by:
CHROMA ATE INC.
Head Office:
43, Wu-Chuan Road, Wu-Ku lnd. Park,
Wu-Ku, Taipei Hsien, Taiwan
TEL: +886-2-2298-3855
FAX: +886-2-2298-3596
http://www.chromaate.com
U. S. A.
7 Chrysler Irvine
CA 92618
TEL: +1-949-421-0355
FAX: +1-949-421-0353
Toll Free: +1-800-478 2026
Distributed by:
Worldwide Distribution and Service Network
19071/19072/19073-2004-02-1000
Europe
Max Planckstraat 4
6716 BE Ede
The Netherlands
Tel: +31-318-648282
Fax: +31-318-648288
China
8F, No.4, Nanyou Tian An
Industrial Estate Shenzhen,
China
Post Code: 518052
Tel: +86-755-
2664-4598
Fax: +86-755-
2641-9620
19071
AC Hipot Tester
19073
AC/DC/IR Hipot Tester
A190701 Remote Control Box
A190702 40kV Test Probe
Production testing of appliances, instruments
and information technology equipment in
accordance with UL, IEC, TUV and other
standards such as EN 60335, EN 60950,
EN 61010, CSA C22.2 No.1010.1, UL 3111
and UL1950
Transformer electrical safety testing
Electric motor safety testing
Variety of electrical components
Testing of electronic components
Specification
Application
Order Information
Model
19071
19073
Mode
ACV
ACV/DCV/IR
Withstanding Voltage Test
Output Voltage
AC : 0.05 ~ 5kV
AC : 0.1 ~ 5kV , DC : 0.1 ~ 6kV
Load Regulation
1 % + 5V
Voltage Resolution
2V
Voltage Accuracy
1% + 5 counts
Cutoff Current
AC : 0.1~20 mA
AC : 0.1~20 mA, DC : 0.01~5 mA
Current Resolution
AC : 1µA, DC : 0.1µA
Current Accuracy
1.5% + 5 counts
Output Frequency
50/60Hz
Test Time
0.1 ~ 999 seconds, continue
Ramp up Time
0.1 ~ 999 seconds, off
Wave Form
Sine wave
Insulation Resistance
Output Voltage
- -
DC : 50 ~ 1000V
Voltage Resolution
- -
2V
Voltage Accuracy
- -
5% + 5 counts
IR Range
- -
0.1MΩ -50GΩ
IR Accuracy
≥ 500V : 1MΩ -1000MΩ± (4% + 5 counts)
1GΩ -10GΩ ± (7% + 5 counts)
- - 10GΩ-50GΩ ± (12% + 5 counts)
< 500V : 0.1MΩ -1000MΩ ± (7% + 5 counts)
< 500V : 0.1MΩ -1000MΩ ±(10% + 5 counts)
Continuity Check
Current 0.1A, 1 Ω ± 0.2Ω, On/Off
Arc Detection
Setting Mode
Programmable Setting
Detection Current
AC : 1mA ~ 15mA, DC : 1mA ~ 5mA
Minimum Pulse Width
10 µs approx
Secure Protection Function
Fast Output Cut-off
0.4 ms after NG happen
Fast DC discharge
0.2 sec.
Ground Fault Interrupt (GFI)
0.5mA ± 0.2mA AC (Low), 5mA ± 2mA AC (High), Close
Panel Operation Lock
Present password
Continuity Check
Current 0.1A, 1Ω ± 0.2Ω, On/Off
GO/ NG Judgment
Indication, Alarm
Go:Short sound ;NG:Long sound
Data Hold
Last tests data memories
Step Hold
Step signal trigger On/Off
Memory Storage
10 tests setups with 60 steps per setup
General
Operation Environment
Temperature : 0 ~ 40˚C
Humidity : ≤ 80% RH
Power Consumption
No load : < 60W
With rated load : 300W max.
Power Requirement
100V/120V/ 220V AC ±10%, 240V AC +5%~ -10%,
50/60Hz
Weight
12kg
Dimension(W X H X D)
270 X 100 X 350mm
AC/DC/IR
.....
Hipot
.....
T
ester
..........................
1. LINE Switch
2. Window Display
3. Reset Button
4. Start Button
5. Function Keys F1~F4
6. Function Key Display Area
7. High Voltage Output Terminal
8. Common Test Terminal
9. Test Indicator
10. Fail Indicator
11. Pass Indicator
12. Calibration Switch
13. Status Line
14. Remote Control
15. LINE Voltage Selector
16. Power Cord Receptacle
17. Earth Ground Terminal
18. Continuity Check
19. Fan
20. Update Switch
Panel Description
1
6
5
4
3
2
16
15
13
17
18
14
19
20