SPECIFICATIONS

INPUT SPECIFICATIONS

VOLTAGE

115 / 230 V selectable, ± 10% variation

FREQUENCY

50/60 Hz ± 5%

FUSE

115 VAC, 230 VAC - 6.3 A Slow-Blo 250 VAC

DIELECTRIC WITHSTAND TEST MODE

OUTPUT RATING

5 KV @ 40 mA AC
5 KV @ 20 mA DC

VOLTAGE DISPLAY

Range:

0.00 - 5.00 KV Full Scale

Resolution: 0.01 KV
Accuracy:

± (2% of reading + 10 V)

HI & LO-LIMIT

AC TOTAL

Range: 0.000 mA - 40.00 mA

AC REAL

Range: 0.000 mA - 40.00 mA

DC

Range: 0.0 - 20,000

µA

CURRENT DISPLAY

AC TOTAL

Range: 0.000 mA - 40.00 mA

AC REAL

Range: 0.000 mA - 40.00 mA

DC

Range: 0.0

µA - 20,000 µA

RAMP HI

Range:

20 mA peak maximum,
ON/OFF selectable

CHARGE LO

Range:

0.000 - 350

µA or Auto Set

DC OUTPUT

≤ 4% Ripple RMS at 5 KV DC @ 20 mA,

RIPPLE

Resistive Load

DISCHARGE TIME

≤ 200 ms

MAXIMUM

1

µF

< 1 KV

0.75

µF < 2 KV

CAPACITIVE LOAD

0.5

µF

< 3 KV

0.08

µF < 4 KV

DC MODE

0.04

µF < 5 KV

AC OUTPUT
WAVEFORM

Sine Wave, Crest Factor = 1.3 - 1.5

OUTPUT

Range:

50/60 Hz, user selectable

FREQUENCY

Accuracy:

± 0.1%

OUTPUT

± (1% of output + 5 V)

REGULATION

from no load to full load and over input
voltage range.

DWELL TIMER

Range:

AC 0.4 - 999.9 sec (0 = Constant)
DC 0.3 - 999.9 sec (0 = Constant)

Accuracy:

± (0.1% + 0.05 sec)

RAMP TIMER

Range: Ramp-Up:

AC 0.1 - 999.9 sec
DC 0.4 - 999.9 sec

Ramp-Down: AC 0.0 - 999.9 sec

DC 0.0,1.0 - 999.9 sec
0.0 = OFF

Accuracy:

± (0.1% + 0.05 sec)

GROUND

Current: DC 0.1 A

± 0.01 A, fixed

CONTINUITY

Max. ground resistance: 1

Ω ± 0.1 Ω, fixed

GROUND FAULT

GFI Trip Current: 450

µA max (AC or DC)

INTERRUPT

HV Shut Down Speed: < 1ms

INSULATION RESISTANCE TEST MODE

VOLTAGE SETTING

Range:

50 - 1000 VDC

Resolution: 1 V
Accuracy:

± (2% of setting + 2 counts)

CHARGING
CURRENT

Maximum >20 mA peak

RESISTANCE

Range: 0.05 M

Ω - 50000 MΩ

DISPLAY

(5 Digit, Auto Ranging)

Accuracy:

50 - 499 V

0.05 M

Ω - 999.9 MΩ,

± (7% of reading +2 counts)

500 - 1000 V

0.05 M

Ω - 999.9 MΩ,

± (2% of reading +2 counts)

1000 M

Ω - 9999 MΩ

± (5% of reading +2 counts)

10000 M

Ω - 50000 MΩ,

± (15% of reading +2 counts)

RAMP HI

Range:

20 mA peak maximum,
ON/OFF selectable

CHARGE LO

Range:

0.000 - 3.500

µA or Auto Set

HI & LO-LIMIT

Range:

0.05 M

Ω - 50,000 MΩ

Accuracy:

Same as Resistance
Display Accuracy

RAMP TIMER

Range:

Ramp-Up:

0.1 - 999.9 sec

Ramp-Down: 1.0-999.9 sec

0 = OFF

Resolution: 0.1 sec
Accuracy:

± (0.1% + 0.05 sec)

DELAY TIMER

Range:

1.0 - 999.9 sec (0 = Constant)

GROUND FAULT

GFI Trip Current: 450

µA max (AC or DC)

INTERRUPT

HV Shut Down Speed: < 1 ms

GROUND BOND TEST MODE

OUTPUT VOLTAGE

Range:

3.00 - 8.00 VAC

(Open Circuit Limit)

Resolution: 0.01 V
Accuracy:

± (2% of setting + 0.03 V)
O.C. Condition

OUTPUT

Range:

50/60 Hz, user selectable

FREQUENCY

Accuracy:

± 0.1%

OUTPUT CURRENT

Range:

1.00 - 30.00 A

Resolution: 0.01 A
Accuracy:

± (2% of setting + 0.02 A)

OUTPUT

Accuracy:

± (1% of output + 0.02 A)

REGULATION

Within maximum load limits, and over
input voltage range

MAXIMUM

1.00 - 9.99 A,

0 - 600 m

LOADING

10.00 - 30.00A,

0 - 150 m

CURRENT DISPLAY

Range:

0.00 - 30.00 A

RESISTANCE
DISPLAY

Range:

0 - 600 m

HI & LO LIMIT

Range:

1.00 - 9.99 A, 0 - 600 m

DWELL TIMER

Range:

0.5 -999.9 sec (0 = Constant)

MILLIOHM OFFSET

Range:

0 - 200 m

CONTINUITY TEST MODE

OUTPUT CURRENT

DC 0.1A

± 0.01A, fixed

RESISTANCE
DISPLAY

Range:

0.00 - 10.00

Ω,

HI & LO-LIMIT

Range:

0.00 - 10.00

DWELL TIMER

Range:

0.0, 0.3 - 999.9
sec (0 = Constant)

MILLIOHM OFFSET

Range:

0.00 - 2.00