Product specifications continued, Dielectric withstand test mode (continued), Ground bond test mode (model 7704 only) – Atec Associated-Research-7704 User Manual
Page 3: General specifications
Product Specifications Continued
Dielectric Withstand Test Mode
(Continued)
Output Frequency
Range: 60 or 50 Hz, user selection
Accuracy: ± 1%
Output 7700/7704
± (1% of output + 5 V) from no load to full load
Regulation
Discharge 7705/7710/
≤ 200 m secs
Time 7715/7720
Dwell Timer 7700/7704
Range: 0, 0.3 - 999.9 sec (0 = Constant)
Resolution: 0.1 sec increments
Accuracy: ± (0.1% + 0.05 sec)
7705/7710/7715/7720
AC Range: 0, 0.3 - 999.9 sec or min (0 = Constant)
DC Range: 0, 0.4 - 999.9 sec or min (0 = Constant)
Resolution: 0.1 second or minute increments
Accuracy: ± (0.1% + 0.05 sec)
Ramp Timer 7700/7704
AC Range: 0.1 - 999.9 sec
DC Range: 0.4 - 999.9 sec
Resolution: 0.1 sec increments
Accuracy: ± (0.1% + 0.05 sec)
7705/7710/7715/7720
Range: 0.1 - 999.9 sec
Resolution: 0.1 sec increments
Accuracy: ± (0.1% + 1 count)
Ground Continuity 7700
Current: DC 0.1 A ± 0.01 A, fixed
Max. Ground Resistance: 1
Ω ± 0.1 Ω, fixed
Ground Fault 7700/7704
GFI Trip Current: 450 µA max (AC or DC)
Interrupt
HV Shut Down Speed: < 1ms
Insulation Resistance Test Mode
(Models 7700 & 7704 only)
Output Voltage
Range: 100 - 1000 V DC
Resolution: 1 V/step
Accuracy: ± (2% of reading + 2 V)
Short Circuit Current
Maximum: 12 mA peak
Voltage Display
Range: 0 - 1000 V
Resolution: 1 V/step
Accuracy: ± (2% of reading + 2 counts)
Resistance Display
Range: 1 - 9999 M
Ω (4 digit, auto ranging)
Resolution: 500 V DC 1000 V DC
M
Ω
M
Ω
M
Ω
0.001
1.000 - 5.388
1.000 - 9.999
0.01
1.40 - 53.88
2.80 - 99.99
0.1
14.0 - 538.8
28.0 - 999.9
1
140 - 9999
280 - 9999
Accuracy: ± (2% of reading + 2 counts) at test voltage
500 - 1000 V and 1 - 1000 M
Ω
± (8% of reading + 2 counts) at test voltage
500 - 1000 V and 1000 - 9999 M
Ω
± (8% of reading + 2 counts) at test voltage
100 - 500 V and 0 - 1000 M
Ω
Charge-LO
Range: 0.000 - 3.500 µA or auto set
HI-Limit
Range: 0 - 9999 M
Ω (0 = OFF)
LO-Limit
Range: 1 - 9999 M
Ω
Delay Timer
Range: 0, 0.5 - 999.9 sec (0 = Constant)
Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec)
Ground Fault Interrupt
GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed: < 1 ms
Ground Bond Test Mode
(Model 7704 only)
Output Voltage
Range: 3.00 - 8.00 V AC
(Open Circuit Limit)
Resolution: 0.01 V/step
Accuracy: ± (2% of setting + 0.03 V) O.C. condition
Output Frequency
Range: 50 or 60 Hz, user selection
Accuracy: ± 1%
Output Current
Range: 3.00 - 30.00 A AC
Resolution: 0.01 A/step
Accuracy: ± (2% of setting + 0.02 A)
Current Display
Range: 0.00 - 30.00 A
Resolution: 0.01 A/step
Accuracy: ± (3% of reading + 0.03 A)
Resistance Display
Range: 0 - 600 m
Ω
Resolution: 1 m
Ω/step
Accuracy: ± (3% of reading + 2 m
Ω)
HI & LO Limit
Range: 0 - 600 m
Ω for 3 - 10 A
0 - 150 m
Ω for 3 - 30 A
Resolution: 1 m
Ω/step
Accuracy: ± (2% of setting + 2 m
Ω)
Dwell Timer
Range: 0, 0.5 - 999.9 sec (0 = constant)
Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec)
Milliohm Offset
Maximum Offset Capability: 200 m
Ω
Resolution: 1 m
Ω/step
Accuracy: ± (2% of setting + 2 m
Ω)
General Specifications
PLC Remote Control
Input: Test, Reset, Recall 1 - 3, Remote Interlock
(Remote Interlock optional on 7700/7704)
Output: Pass, Fail, Test-in-Process
Interface
Standard RS-232, Optional GPIB
Memory 7700/7704
50 memories w/8 Steps per memor y
7705/7710/7715/7720
50 Memories
Security
Programmable password lockout capability to avoid
unauthorized access to test set-up program
Safety
Built-in SmartGFI
®
circuit
Display
2 x 20 characters with front panel contrast setting
Alarm Volume Setting
Front panel adjustable with 10 set points
Line Cord
Detachable 7 ft. (2.13 m) power cable terminated in
a three prong grounding plug
Terminations
Detachable 5 ft. (1.52 m) high voltage & return
lead with clips
Mechanical
Tilt up front feet
Dimensions 7700/7704
(WxHxD) 17 x 5.8 x 16.5 in. (432 x 147 x 419 mm)
7705/7710/7715/7720
(WxHxD) 17 x 5.8 x 15.75 in. (432 x 147 x 400 mm)
Weight 7700
61.65 lbs (28 kgs)
7704
68.75 lbs (31.25 kgs)
7705/7710/7715/7720
48.7 lbs (22.1 kgs)
Environmental
Operating temperature: 32˚ - 104˚ F (0˚ - 40˚ C)
Relative humidity: 20% - 80%
Calibration
Traceable to National Institute of Standards &
Technology (NIST). Calibration controlled by software.
Adjustments are made through front panel keypad in
a restricted access calibration mode. Calibration
information stored in non-volatile memor y.