Product specifications, Input specifications, Dielectric withstand test mode – Atec Associated-Research-7704 User Manual
Page 2: Dielectric withstand test mode (continued)
Product Specifications
Input Specifications
Voltage 7700/7704
100/115/200/230 VAC ± 10%,
single phase, user selection
7705/7710/7715/7720
115/230 VAC ± 10%, single phase, user selection
Frequency 7700/7704
47 - 63 Hz
7705/7710/7715/7720
50/60 Hz ± 5%
Fuse 7700/7704
15 Amp 250 V fast blow internal
7705/7710/7715/7720
10 Amp, 250 V
Dielectric Withstand Test Mode
Output 7700/7704
5 KV @ 100 mA AC, 6 KV @ 10 mA DC
Rating
7705
10 KV @ 20 mA AC
7710
12 KV @ 10 mA DC
7715
20 KV @ 10 mA AC
7720
20 KV @ 5 mA DC
Output
7700/7704
Range: 0 - 5 KV AC, 0 - 6 KV DC
Adjustment
Resolution: 1 V/step
Accuracy: ± (2% of setting + 5 V)
7705
Range: 0 - 10 KV AC
Resolution: 1V
Accuracy: ± (2% of setting + 10 V)
7710
Range: 0 - 12 KV DC
Resolution: 1V
Accuracy: ± (2% of setting + 10 V)
7715
Range: 0 - 20 KV AC
Resolution: 10 V/step
Accuracy: ± (2% of setting + 10 V)
7720
Range: 0 - 20 KV DC
Resolution: 10 V/step
Accuracy: ± (2% of setting + 10 V)
HI-Limit 7700/7704
AC Range: 0.00 - 99.00 mA
and LO-Limit
Resolution: 0.01 mA/step
DC Range: 0 - 9999 µA
Resolution: 1 µA/step
Accuracy: AC or DC ± (2% of setting + 2 counts)
7705
Range 1: 0.0 - 9.999 mA
Resolution: 0.001 mA
Range 2: 10.00 - 20.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of setting + 2 counts)
7710
Range 1: 0.000 - 999.9 µA
Resolution: 0.1 µA
Range 2: 1000 - 9999 µA
Resolution: 1 µA
Accuracy: ± (2% of setting + 2 counts)
7715
Range: 0.00 - 10.00 mA
Resolution: 0.01 mA/step
Accuracy: ± (2% of setting + 2 counts)
7720
Range: 0.0 - 5000 µA
Resolution: 1 µA
Accuracy: ± (2% of setting + 2 counts)
DC Ramp HI 7700/7704
12 mA peak maximum, (ON/OFF selectable all models)
DC Charge LO7700/7704
Range: 0.0 - 350 µA DC or auto set
Arc Detection
Range: 1 - 9
Failure Detector
Audible and visual
Dielectric Withstand Test Mode
(Continued)
Voltage
7700/7704
Range: 0.00 - 6.00 KV full scale
Display
Resolution: 10 V/step
Accuracy: ± (2% of reading + 2 counts)
7705
Range: 0.00 - 10.00 KV Full scale
Resolution: 0.01 mA
Accuracy: ± (2% of reading + 20 V)
7710
Range: 0.00 - 12.00 KV Full scale
Resolution: 0.01 KV
Accuracy: ± (2% of reading + 2 counts)
7715
Range: 0.00 - 20.00 KV Full scale
Resolution: 10 V
Accuracy: ± (2% of reading + 20 V)
7720
Range: 0.00 - 20.00 KV Full scale
Resolution: 10 V
Accuracy: ± (2% of reading + 20 V)
Current 7700/7704
Auto Range
Display
AC
Range 1: 0.000 mA - 3.500 mA
Resolution: 0.001 mA/step
Accuracy: ± (2% of reading + 0.003 mA)
Range 2: 3.00 - 99.00 mA
Resolution: 0.01 mA/step
Accuracy: ± (2% of reading + 0.06 mA)
DC
Range 0.0 µA - 350.0 µA
Resolution: 0.1 µA/step
Accuracy: ± (2% of reading + 0.3 µA)
Range 2: 300 µA - 3500 µA
Resolution: 1 µA/step
Accuracy: ± (2% of reading + 2 µA)
Range 3: 3000 µA - 9990 µA
Resolution: 10 µA/step
Accuracy: ± (2% of reading + 60 µA)
7705
Auto Range
Range 1: 0.000 mA - 3.500 mA
Resolution: 0.001 mA
Range 2: 3.00 - 20.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of reading + 3 counts)
7710
Auto Range
Range 1: 0.0 - 350.0 µA
Resolution: 0.001 µA
Range 2: 300 - 5000 µA
Resolution: 0.01 µA
Accuracy: ± (2% of reading + 3 counts)
7715
Auto Range
Range 1: 0.000 mA - 3.500 mA
Resolution: 0.001 mA
Range 2: 3.00 - 10.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of reading + 3 counts)
7720
Auto Range
Range 1: 0.0 - 350.0 µA
Resolution: 0.1 µA
Range 2: 300 - 5000 µA
Resolution: 1 µA
Accuracy: ± (2% of reading + 3 counts)
DC Output
7700/7704
≤ 4% Ripple RMS at 6 KV DC @ 3.5 mA, Resistive load
Ripple 7710
< 5% (12 KV/9999 µA at Resistive Load)
7720
< 5% (20 KV/4999 µA at Resistive Load)
AC Output Waveform
Sine Wave, Crest Factor = 1.3 - 1.5
AC Output 7705/7710/
± (1% of setting + 10 V) from no load to full load
Regulation 7715/7720