Noise figure analyzers, Nfa series key specifications – Atec Agilent-N8972A_N8973A_N8974A_N8975A User Manual
Page 2
10 MHz to 1.5 GHz, 3 GHz, 6.7 GHz and 26.5 GHz Noise Figure Analyzers (cont.)
Noise Figure Analyzers
N8972A
N8973A
N8974A
N8975A
NFA Series Key Specifications
Specifications apply over 0°C to +55°C unless otherwise noted. The
analyzer will meet its specifications after 2 hours of storage within
the operating temperature range, 60 minutes after the analyzer is
turned on, with Alignment running. A user calibration is required
before corrected measurements can be made.
Frequency Range
NFA Series:
N8972A
10 MHz to 1.5 GHz
N8973A
10 MHz to 3 GHz
N8974A
10 MHz to 6.7 GHz
N8975A
10 MHz to 26.5 GHz
Measurement Speed (nominal)
8 Averages
64 Averages
N8972A:
<100 ms/measurement
<80 ms/measurement
N8973A:
<50 ms/measurement
<42 ms/measurement
N8974A:
<70 ms/measurement
<50 ms/measurement
N8975A:
<70 ms/measurement
<50 ms/measurement
Measurement Bandwidth (nominal)
N8972A: 4 MHz
N8973A, N8974A, N8975A: 4 MHz, 2 MHz, 1 MHz, 400 kHz, 200 kHz, 100 kHz
Noise Figure and Gain
(Performance is dependent upon ENR of noise source used)
N8972A
Noise Source ENR
4 – 7 dB
12 – 17 dB
20 – 22 dB
Noise Figure
Measurement range
0 to 20 dB
0 to 30 dB
0 to 35 dB
Instrument uncertainty
±<0.1 dB
±<0.1 dB
±<0.15 dB
Gain
Measurement range
–20 to +40 dB
Instrument uncertainty
±<0.17 dB
N8973A, N8974A and
Noise Source ENR
N8975A
(10 MHz to 3.0 GHz)
4 – 7 dB
12 – 17 dB
20 – 22 dB
Noise Figure
Measurement range
0 to 20 dB
0 to 30 dB
0 to 35 dB
Instrument uncertainty
±<0.05 dB
±<0.05 dB
±<0.1 dB
Gain
Measurement range
–20 to +40 dB
Instrument uncertainty
±<0.17 dB
N8974A and N8975A
Noise Source ENR
(>3.0 GHz)
4 – 7 dB
12 – 17 dB
20 – 22 dB
Noise Figure
Measurement range
0 to 20 dB
0 to 30 dB
0 to 35 dB
Instrument uncertainty
±<0.15 dB
±<0.15 dB
±<0.2 dB
Gain
Measurement range
–20 to +40 dB
Instrument uncertainty
±<0.17 dB
Characteristic
1
Noise figure at 23ºC ± 3ºC (10 MHz to 3.0 GHz)
Characteristic
1
Noise figure at 23ºC ± 3ºC (3.0 GHz to 26.5 GHz)
Characteristic values are met or bettered by 90% of instruments with 90%
confidence.
Frequency Reference
Standard
Opt.1D5
Aging
±<2 ppm
1
/year
±<0.1 ppm/year
Temperature stability
±<6 ppm
±<0.01 ppm
Settability
±<0.5 ppm
±<0.01 ppm
Tuning Accuracy (Start, Stop, Center, Marker)
4 MHz Measurement Bandwidth (default on all models of Noise Figure
Analyzer)
Frequency
Error
10 MHz – 3.0 GHz
± 3.0 GHz – 26.5 GHz ± <4MHz Measurement Bandwidth (functionality not present in N8972A) Error 10 MHz – 3.0 GHz ± 3.0 GHz – 26.5 GHz ± bandwidth 1 Parts Per Million (10e-6) 10 8 4 0 10 500 1000 1500 2000 2500 3000 Frequency (MHz) Noise Figure (dB) 89 10 11 12 Frequency (MHz) Noise Figure (dB) 3000 3883 4707 5651 7418 8303 8187 9187 10072 10956 11810 12724 13608 14492 15376 16260 17145 18029 18913 18797 10681 21565 22419 23334 24210 25102 25986 26500
Frequency
6
2