2700 series dual domain audio analyzer – Atec Audio-Precision-ats1dd User Manual
Page 2
2
a p . c o m
O p t i o n s
“A” models include AP2700 control software, APIB or USB interface, cable and full set of user
manuals. “G” models include IEEE-488.2 GPIB control interface, cables and manuals.
Analog system 1 kHz THD+N, 20 kHz BW ≤ –112 dB
AP’s flagship analyzer, the 2700 Series is designed for audio
engineers who need the highest performance, lowest distortion
and greatest flexibility possible in their audio analyzer.
The 2700 Series has the best performance of any audio
analyzer in the world: its high-precision analog audio design is
markedly superior to the best digital implementations in several
key areas, especially residual noise and distortion, and high-
bandwidth.
The unparalleled accuracy of a dedicated hardware instrument
Fast instrument operation and powerful analysis under
sophisticated control software
Programmatic control for high-speed automation
Serial digital interface testing
True dual domain instrument for highest performance
192 kHz digital input and output capabilities
500 kHz analog / 100 kHz digital measurement bandwidth
Vanishingly low residual noise and THD+N
Generates and analyzes a broad palette of stimulus signals such as
MLS, white and pink noise and other special purpose waveforms
Connects to control PC via USB
•
•
•
•
•
•
•
•
•
•
2700 Series Dual Domain Audio Analyzer
2 7 0 0 S e r i e s K e y S p e c i f i c a t i o n s
SYS-2722A/G
Dual domain. Two-channel analog input/output, analog analyzer and
generator, DSP analyzer and generator, plus 192k digital input/output
SYS-2722A-M
Windows Vista Logo Test System: SYS-2722A with all required filters
and cables for performing Windows Vista Logo audio tests
SYS-2720A/G
Digital domain only. DSP analyzer and generator with 192k digital
input/output
SYS-2712A/G
Analog domain only. Two-channel analog input/output, analog analyzer
and generator, DSP analyzer and generator
SYS-2702A/G
Analog domain only. Two-channel analog input/output, analog analyzer
and generator
2700-PROG
Programmer’s Starter Kit, including: GPIB Programmer’s Reference
Manual and GPIB utilities and samples CD-ROM
S2-IMD
Adds IMD generator & analyzer to SYS-2702, SYS-2712, or SYS-2722
S2-W&F
Adds wow and flutter analyzer to SYS-2702, SYS-2712, or SYS-2722
S2-BUR
Adds tone burst, square wave and noise generator to SYS-2702, SYS-
2712, or SYS-2722
S-AES17
Adds multistage low-pass filter for D/A measurements to AES17
standard. Includes special pre-analyzer low-pass filter and analyzer
low-pass option filters FLP-B20k and FLP-B40k
OPT-2020
Adds special pre-analyzer low-pass filter for measurements of signals
with large amounts of out-of-band noise
OPT-2711
Adds the Dolby Digital Generator option to a new SYS-2722 or SYS-
2720, allowing the encoding of stimulus tones in Dolby Digital (AC-3)
before digital output. (Not compatible with GPIB instrument)
EWP2-2700
Two-year extended warranty for 2700 Series
BASIC FORMAT
Channels
2
Computer interface
USB, APIB or GPIB
OS Compatability
Windows Vista, XP
Dimensions
3U
GENERATOR PERFORMANCE
Sine Frequency Range
10 Hz–204 kHz
Frequency Accuracy
2 ppm
IMD Test Signals
SMPTE, CCIF, DFD, DIM, TIM, DIN,
Maximum Amplitude (balanced)
26.43 Vrms
Amplitude Accuracy
±0.06 dB
Flatness (20 Hz–20 kHz)
±0.008 dB (typically <0.003 dB)
Residual THD+N, 1 kHz
–112 dB, + 1.0 μV , 22 kHz BW
Analog Output Configurations
unbalanced & balanced
Digital Output Sampling Rate
28 kHz–200 kHz
Dolby Digital Generator
Yes
ANALYZER PERFORMANCE
Maximum Rated Input Voltage
230 Vpk, 160 Vrms (dc to 20 kHz),
overload protected in all ranges
Maximum Bandwidth
>500 kHz
Amplitude Accuracy (1 kHz)
±0.05 dB
Amplitude Flatness (20 Hz–20 kHz)
±0.02 dB
Residual Input Noise (22 kHz BW)
≤
1.0 μV [–117.8 dBu]
Residual THD+N (20 kHz BW)
–112 dB + 1.0 µV, 22 kHz BW
FFT Resolution
Up to 4 M samples (87 seconds at a
sample rate of 48 kHz)
IMD Measurement Capability
SMPTE, CCIF, DFD, DIM, TIM, DIN
DC Voltage Measurement
Yes, with DCX-127 Multifunction Module