beautypg.com

Internal ocxo timebase, Synchronization inputs/outputs, General specifi cations – Atec Agilent-E6601A User Manual

Page 5: External reference input, Baseband triggers

background image

5

Internal OCXO Timebase

Output level range

Typically 0 to +10 dBm

Aging rate <

±0.1

ppm/year

Temperature stability (referenced to +25 °C)
–10 to +70 °C

< ±0.05 ppm

Accuracy
After 30-minute warm-up

±[time since last calibration x

aging rate + temperature stability

+ accuracy of calibration]

Initial adjustment ±0.05

ppm

Locking range

Typically ±0.2 ppm

Output frequency

10 MHz + locking range

Output impedance

Typically 50 ohms

Synchronization Inputs/Outputs

External reference input

Locking range

Typically ±0.2 ppm

Input frequency

1, 2, 5, or 10 MHz ± locking range

Input level range

Typically 0 to +10 dBm

Input impedance

Typically 50 ohms

Baseband triggers

Rear-panel input and/or output connections for format-dependent
synchronization with external equipment.

Bi-directional
DB9 connector

5 TTL triggers

Input
BNC connector

1 TTL trigger input

Output
BNC connector

1 TTL trigger output

General Specifi cations

Operating temperature range

+10 to +55 °C

Storage temperature range

–20 to +65 °C

Dimensions H x W x D

8.75 x 16.75 x 21 inches

222 x 426 x 533 mm

Weight 51

pounds

23.1

kg

AC power input

100 to 240 VAC,

50 to 60 Hz

AC power consumption

Typically 260 W maximum

Calibration interval 2

years

Self alignment conditions

User alerted that automatic
self re-alignment is required
and can choose to align
then or to ignore until later

> ±10 °C from last self align-
ment, or > 24 hours since last
RF generator self alignment or
> 30 days since last instrument
self alignment

Self alignment times
RF

generator

Typically < 1 minute

Instrument and RF generator Typically < 5 minutes

EMC
Meets standards as listed

IEC 61326:2002 /

EN 61326:1997 + A1:1998 +

A2:2000 + A3:2003

Canada

ICES-001:2004

Australia and New Zealand

AS/NZS CISPR11:2002

Safety
Meets standards as listed

IEC 61010-1:2001 /

EN

61010-1:2001

Canada

CSA C22.2 No. 61010-1:2004

USA

UL

61010-1:2004

Radiated source leakage

Within cellular frequency
bands, at zero span with
10 Hz resolution bandwidth
and one inch (2.54 cm) from
instrument front panel and
from front half of all sides of
instrument surfaces

Typically < 1 µV rms

Remote programming
GPIB

IEEE

standard

488.2

LAN

1 RJ45 rear-panel connector

USB-B

1

rear-panel

connector

External device connections
USB-A

2 front-panel USB 1.1

4 rear-panel USB 2.0