Rf analyzer, Rf channel suite, Channel power measurement – Atec Agilent-E6601A User Manual
Page 3: Frequency error measurement
![background image](/manuals/317981/3/background.png)
3
RF Analyzer
Frequency ranges
Cellular bands
411 to 486 MHz
776 to 960 MHz
1574 to 1577 MHz
1710 to 1980 MHz
General purpose
400 to 2000 MHz
Input level ranges
Average power
–65 to +33 dBm, typical
over-range to +35 dBm
Peak power
–65 to +37 dBm (5 W)
Self alignment validity
≤ ±10 °C change and ≤ 30 days
from last self alignment
VSWR
RF
IN/OUT
400 to 1000 MHz
< 1.15:1
1000 to 2000 MHz
< 1.2:1
2000 to 2200 MHz
< 1.3:1
2200 to 2600 MHz
< 1.35:1
Filters
1 kHz
30
kHz
100
kHz
300
kHz
640
kHz
1.23
MHz
1.6
MHz
3.84
MHz
5.0
MHz
GSM Tx power
W-CDMA mean power
Trigger setup
Arm
Single,
continuous
Delay (varies by fi lter
–180 to 180 ms
and measurement)
Sources
External, fall, immediate, rise
(varies by measurement)
Measurement setup
Averaging
Off, 1 to 999
(multi-measurement
count–not applicable to
all measurements)
Timeout
Off, 0.1 to 999.9 s
RF Channel Suite
Includes channel power, frequency error, power versus time, and
IQ capture measurements.
Input level range
Average power
–65 to +35 dBm
Measurement interval and fi lter ranges
1
kHz
fi lter
1 to 5000 ms
30 kHz fi lter
0.4 to 4000 ms
100 kHz fi lter
0.1 to 4000 ms
300 kHz fi lter
0.01 to 1700 ms
640 kHz fi lter
0.01 to 810 ms
1.23 MHz fi lter
0.01 to 420 ms
1.6 MHz fi lter
0.01 to 320 ms
3.84 MHz fi lter
0.01 to 135 ms
5.0 MHz fi lter
0.01 to 100 ms
GSM Tx power fi lter
0.01 to 199 ms
W-CDMA mean power fi lter 0.1 to 3200 ms
Channel power measurement
Measurement accuracy¹
Within cellular frequency bands
–59 to +35 dBm
< ±0.6 dB, typically < ±0.3 dB
–65 to < –59 dBm
< ±0.7 dB, typically < ±0.4 dB
Within cellular frequency bands with < 48 hours warm-up before
self alignment initiated
–59 to +35 dBm
< ±0.7 dB, typically < ±0.3 dB
–65 to < –59 dBm
< ±0.8 dB, typically < ±0.4 dB
400 to 2000 MHz
–59 to +35 dBm
Typically < ±0.4 dB
–65 to < –59 dBm
Typically < ±0.6 dB
Measurement repeatability
Typically < ±0.05 dB
Returning to same level and frequency, no temperature change
and insignifi cant time change
Frequency error measurement
Measurement accuracy
CW signals from 400 to 2000 MHz
–60 to +35 dBm
< ±(50 Hz + timebase accuracy)
with 1, 30, 100,
300, 640 kHz GSM
Tx power fi lters
–40 to +35 dBm
< ±(50 Hz + timebase accuracy)
with 1.23, 3.84,
5.0 MHz W-CDMA
mean power fi lters
1. Additional accuracy error when using RF OUT ONLY port is < ±0.1 dB.