beautypg.com

Ag ilent e6006a power meter submodule – Atec Agilent-E6008B User Manual

Page 6

background image

6

Agilent E6006A Power Meter
Submodule

2

Characteristics

Sensor element: InGaAs

Wavelength range: 800 – 1650 nm

Calibrated wavelengths: 850 nm,
1300 nm, 1310 nm, 1550 nm, 1625 nm
(special wavelengths upon request).

Power range: +10 to –70 dBm

Max. input power (damage level):
+13 dBm / 20 mW

Display Resolution: 0.01 dB

Display Units: dBm, dB, mW,

µW, nW,

pW

Display Contents

• Calibrated λ in nm

• Modulation frequency in Hz

• Reference value in dB

Display Updates per second: 3

Optical input: User-exchangeable
Connector Interface

Applicable fiber type: 9/125

µm,

50/125

µm, 62.5/125 µm

2

E6006A Power Meter Submodule not

available for E6013A 1310/1550/1625 nm
Mini-OTDR module

Specifications

Uncertainty at reference conditions:
± 3%

Power level: –20 dBm

Continuous Wave (CW)

Wavelength: 1300 nm

± 3 nm,

1310 nm

± 3 nm, 1550 nm ± 3 nm

Fiber type: 50/125

µm graded index,

Agilent/HMS-10 connector

Spectral bandwidth: up to 10 nm

Ambient temperature: +18°C to + 28°C

At day of calibration (add 0.3% for aging
of over one year; add 0.6% for aging of
over two years).

Total uncertainty:

± 5% ± 0.5 nW

(1310, 1550 nm)

Power level: +0 to –50 dBm

Continuous Wave (CW)

Wavelength: 850 nm

± 3 nm,

1300 nm

± 3 nm, 1310 nm ± 3 nm,

1550 nm

± 3 nm,

Fiber type: SM to 50

µm graded index

(add 2% to total uncertainty for fiber
62.5

µm).

Straight and angled connectors

Ambient temperature: +10°C to +40°C

Within 2 years after calibration

Supplementary Performance
Characteristics

• Automatic Zeroing Circuitry

• Automatic Ranging

• Modulation frequency recognition

(270 Hz, 1 kHz, 2 kHz) is available at
power level between
+10 and –45dBm (peak amplitude).

• Dual Wavelength measurement is

available at power levels between
+10 and –45 dBm (peak amplitude).

• Wavelength encoding recognition

(350 Hz, 550 Hz) is available at power
levels between +10 and –45 dBm
(peak amplitude).

• Reference value is presettable from

+30 to –80 dBm

• Each calibrated wavelength has its

own reference memory.

• The actual display content can be

transferred to reference memory
(DISP

→REF).

• Hold Data functionality

General Specifications

Dimensions: ca. 120 mm H x 40 mm W
x 25 mm D (4.7” x 1.6” x 1.0”)

Weight: <130 g.

Operating Temperature: 0°C to +50 °C

Storage Temperature: –40°C to +60°C

Humidity: 95% R.H. from 0°C to 40°C
non cond.

Recommended Recalibration Period:
2 years