beautypg.com

Atec Agilent-8156A User Manual

Page 5

background image

Page 5 of 8

Variable Optical Attenuator Modules with Power Control (Single Mode Fibers)

81576A 81577A

Connectivity

straight connector, versatile interface

angled connector, versatile interface

Fiber type

9/125

µm SMF

Wavelength range

1250 nm – 1650 nm

Attenuation range

0 – 60 dB

Resolution 0.001

dB

Attenuation Setting

Power Setting

Attenuation Setting

Power Setting

Repeatability

± 0.010 dB

± 0.015 dB

[2]

± 0.010 dB

± 0.015 dB

[2]

Accuracy (uncertainty)

[1], [3], [4], [5]

± 0.1 dB

± 0.1 dB

Settling time

[6]

typ. 100 ms

typ. 300 ms

typ. 100 ms

typ. 300 ms

Transition speed

typ. 0.1 – 12 dB/s

Relative power meter uncertainty

[7], [8]

± 0.03 dB ± 200 pW

Attenuation flatness

[1], [5], [9], [10]

± 0.07 dB; typ. ± 0.05 dB (1520 nm < λ < 1620 nm)

typ.

± 0.10 dB (1450 nm < λ < 1640 nm)

Spectral ripple

[11]

typ.

± 0.003 dB

Insertion loss

[3], [5], [12], [13]

typ. 0.9 dB (excluding connectors)

1.8 dB; typ. 1.2 dB (including connectors)

[14]

Insertion loss flatness

[1], [14]

typ.

± 0.1 dB (1420 nm < λ < 1615 nm)

[5]

Polarization dependent loss

[3], [12], [14]

0.10 dBpp; typ. 0.05 dBpp

Return loss

[12], [14]

typ. 45 dB

typ. 57 dB

Maximum input power

[15]

+ 33 dBm

Shutter isolation

typ. 100 dB

Dimensions (H x W x D)

75 mm x 64 mm x 335 mm (2.8” x 2.6” x 13.2”)

Weight 1.3

kg

Recommended recalibration period

2 years

Operating temperature

10°C – 45°C

Humidity Non-condensing
Warm-up time

30 minutes

[1]

At constant operating temperature

[2]

Output power > -40 dBm, input power < +27 dBm. For input power > +27 dBm add typ.

± 0.01 dB

[3]

Temperature within 23°C

± 5°C

[4]

Input Power < +30 dBm;

λ = 1550 nm ± 15 nm; typ. for 1250 nm < λ < 1650 nm

[5]

For unpolarized light

[6]

Step size < 1 dB; for full range: typ. 6 seconds

[7]

Wavelength and SOP constant; temperature constant and between 23°C

± 5°C; λ < 1630 nm; 1s averaging time

[8]

Output power >-45dBm,

input power

≤ 27dBm, for input power > + 27dBm add typ. ± 0.02dB

[9]

Relative to reference at 0 dB attenuation

[10]

λ

disp

set to 1550 nm; attenuation

≤ 20 dB;

for attenuation > 20 dB:
add typ. 0.01 dB (

α [dB] – 20) for 1520 nm < λ < 1620 nm

add typ. 0.02 dB (

α [dB] – 20) for 1450 nm < λ < 1640 nm

[11]

Linewidth of source

≥ 100 MHz

[12]

For

λ = 1550 nm ± 15 nm

[13]

Add typ. 0.1 dB for

λ = 1310 nm ± 15 nm

[14]

Measured with Agilent reference connectors

[15]

Agilent Technologies Deutschland GmbH assumes no responsibility for damages caused by scratched or poorly cleaned connectors.