Commissioning
ProLab 4000
30
ba75646e02
01/2010
3.7.2
USB-B interface (USB Device)
You can connect a PC to the meter via the USB-B interface(USB
Device).
Data that is transmitted to the interface with
out with a printer or received with a terminal program if a PC is con-
nected.
The USB driver of the enclosed CD-ROM must be installed for the inter-
face to be available for the PC.
Installation of the USB
driver on the PC
System requirements of the PC for installation of the USB driver:
z PC with Pentium processor or higher with at least one free USB con-
nection and CD-ROM drive
z Windows 2000, XP.
Operation at USB-B
(USB Device)
Connect the interface on the ProLab 4000 to the PC using a cable with
a USB-A and a USB-B plug.
Set up the following transmission data on the PC/printer:
1
Insert the supplied installation CD in the CD drive of your PC.
2
Connect the meter to the USB input of the PC via the USB
cable.
Windows automatically installs the driver for the meter. Follow
the Windows installation instructions as necessary.
The meter is listed as a virtual COM interface among the con-
nections in the Windows instrument manager.
Baud rate
selectable between: 1200 ... 115200
The baud rate must agree with the baud rate set on
the PC/printer.
Handshake
RTS/CTS
PC only:
Parity
none
Data bits
8
Stop bits
1