7 selecting and adjusting the nonlinear, Temperature compensation tc – Xylem ProLab 4000 MULTIPARAMETER BENCHTOP METER User Manual
Page 140

Conductivity
ProLab 4000
140
ba75646d02
01/2010
10.2.7
Selecting and adjusting the nonlinear
temperature compensation TC nLin1
The nonlinear temperature compensation is carried out according to
the following equation:
The equation includes two temperature coefficients, TC
1
and TC
2
. In
the ProLab 4000 meter, the temperature coefficients TC
1
and TC
2
are
stored for the following test samples:
Suitable temperature coefficients TC
1
and TC
2
for solutions that are not
mentioned here can be taken from tables found in further reading and
then entered in the meter.
X
Tref
= X
T
/ (1 + TC
1
·[T-TRef] + TC
2
·[T-TRef]
2
)
X
T
= Conductivity at temperature T
[S/cm]
X
Tref
= Conductivity at reference temperature Tref
[S/cm]
TC
1
= Temperature
coefficient
1
[1/°C]
TC
2
= Temperature
coefficient
2
[1/°C
2
]
T
= Temperature
[°C]
Tref
= Reference temperature
[°C]
HCl
NaOH
NaCl
KCl
Concentration c
[mol/l]
1,0
0,67
0,85
0,67
Temperature range
[°C]
0 … 55
0 … 40
0 … 40
0 … 40
Tref [°C]
25
25
25
25
1
Highlight the channel of the conductivity measuring cell in the
measured value display.
2
If necessary, select the measured parameter with
3
Open the input window for the temperature coefficients with
Calibration / TC calibration / TC nLin1....