Page 84: Ph/orp

Technical data
ProLab 1000
84
ba75576e04
01/2011
Stand March 3, 2010
Translation of the legally binding German version
7.2
Measuring ranges, resolution, accuracy
7.2.1
pH/ORP
Note
The measuring ranges and accuracy values specified here apply
exclusively to the meter. The accuracy of the sensors has also to be
taken into account.
Instrument safety
EC guideline 2006/95/EC
EN 61010-1
ANSI/UL 61010-1
CAN/CSA-C22.2 No. 61010-1
Radio data
transmission
EC guideline 1999/5/EC
EN 300 330-2
EN 50364
EN 60950-1
Climatic class
VDI/VDE 3540
IP protection class
EN 60529
Measuring ranges,
resolution
Variable
Measuring range
Resolution
pH
- 2.000 ... + 20.000
- 2.00 ... + 20.00
0.001
0.01
U [mV]
- 1999.9 ... + 1999.9
- 1999 ... + 1999
0.1
1
T [°C]
- 10.0 ... + 120.0
0.1
Manual
temperature input
Variable
Range
Increment
T
manual
[°C]
- 10 ... + 130
1
Accuracy (± 1 digit)
Variable
Accuracy
Temperature of the test
sample
pH
± 0.003
± 0.01
+ 15 °C ... + 35 °C
U [mV] / range
- 999.9 ... + 999.9
± 0.2
+ 15 °C ... + 35 °C
- 1999 ... + 1999
± 1
+ 15 °C ... + 35 °C
T [°C] / temperature sensor
± 0.1
0 °C ... + 55 °C