4 calibrating for ise measurements, Calibrating for ise measurements – Xylem ProLab 1000 BENCH TOP pH METER User Manual
Page 55

ProLab 1000
Operation
55
ba75576e04
01/2011
4.6.4
Calibrating for ISE measurements
Why calibrate?
Ion-selective electrodes age and are temperature-dependent. This
changes the slope. As a result, an inexact measured value is displayed.
Calibration determines the current value of the slope of the electrode
and stores it in the instrument.
Thus, you should calibrate before each measurement and at regular
intervals.
When to calibrate?
z Before any measurement if possible
z After connecting another ISE electrode
z When the CalClock flashes, e. g. after a voltage interruption
Standard solutions
Use two to seven different standard solutions. The standard solutions
have to be selected in either increasing or decreasing order.
Note
The unit of the standard solution and measurement result is selected in
the Calibration / Unit menu.
Menu item
Possible
setting
Description
Calibration /
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration /
Calibration history
-
Displays the calibration
history of the last calibrations.
Man. temperature
-20 ... +130 °C
Entry of the manually
determined temperature. For
measurements without
temperature sensor only.
Ion type
Ag, Br, Ca, Cd,
Cl, CN, Cu, F,
I, K, Na, NO3,
Pb, S, NH3,
NH4, CO2
Selection of the ion type to be
measured
Unit
mol/l, mg/kg,
ppm, %, mg/l
Selection of the unit for the
display of the measurement
result and calibration
standards.
Density
0.001 ... 9.999
g/ml, kg/l
Adjustable density of the test
sample (only with
Unit: mg/kg, ppm, %)