What to do if...

47

6

What to do if...

Error message

0FL

0FL

0FL

0FL

,

UFL

UFL

UFL

UFL

Error message

Err3

Err3

Err3

Err3

Error message

Err4

Err4

Err4

Err4

Error message

Err5

Err5

Err5

Err5

Cause

Remedy

– The measured value lies outside

the measuring range

– Sensor module defective

– Replace sensor module

Cause

Remedy

– Erroneous calibration

– The data of the sensor module is

outside the specified characteris-
tics.

– Recalibrate

– Reset to default setting (see

section 4.12)

– Sensor module contaminated

– Clean and replace sensor

module if necessary

– Moisture in the plug connection

– Dry plug and socket

– Calibration solution not suitable

– Check calibration solution

Cause

Remedy

No sensor module

– Connect sensor module

– Check the plug connection

– Switch the meter off and then on

again.

– Perform a processor reset (see

section 6.1)

– Replace sensor module

Cause

Remedy

Sensor module defective
or
Wrong sensor module

– Check the plug connection

– Switch the meter off and then on

again.

– Perform a processor reset (see

section 6.1)

– Replace sensor module