6 what to do if, What to do if, Ee chapter 6 w – Xylem VARIO COND TEMPERATURE CONDUCTIVITY METERS User Manual
Page 47: Chapter 6 w, 6what to do if
VARIO Cond
What to do if...
47
6
What to do if...
Error message
0FL
0FL
0FL
0FL
,
UFL
UFL
UFL
UFL
Error message
Err3
Err3
Err3
Err3
Error message
Err4
Err4
Err4
Err4
Error message
Err5
Err5
Err5
Err5
Cause
Remedy
– The measured value lies outside
the measuring range
– Sensor module defective
– Replace sensor module
Cause
Remedy
– Erroneous calibration
– The data of the sensor module is
outside the specified characteris-
tics.
– Recalibrate
– Reset to default setting (see
– Sensor module contaminated
– Clean and replace sensor
module if necessary
– Moisture in the plug connection
– Dry plug and socket
– Calibration solution not suitable
– Check calibration solution
Cause
Remedy
No sensor module
– Connect sensor module
– Check the plug connection
– Switch the meter off and then on
again.
– Perform a processor reset (see
– Replace sensor module
Cause
Remedy
Sensor module defective
or
Wrong sensor module
– Check the plug connection
– Switch the meter off and then on
again.
– Perform a processor reset (see
– Replace sensor module