A l c – Xylem VARIO COND TEMPERATURE CONDUCTIVITY METERS User Manual
Page 24

Operation
VARIO Cond
24
4.4.2
Determining the cell constant (calibration in the control stan-
dard)
Note
This method of automatically determining the cell constant by calibration in
the 0.01 mol/l KCL control standard solution can only be used for measuring
cells with cell constants in the range 0.450 ... 0.500 cm
-1
(sensor module
TetraCon V).
This is how you can determine the cell constant:
Note
You can return to the measuring mode with <
g
>.
The displayed calibrated cell constant is used immediately.
Note
You can terminate the calibration with <
E
>.
The data of the terminated calibration is not used.
1
Open the setting menu for the cell constant with <
C
>.
The display shows the setting last selected.
2
If necessary, press <
"_
> repeatedly until the CAL display appears.
3
If necessary, switch between the cell constant last calibrated, the
date of the calibration and the calibration temperature with <
§
>.
4
To calibrate:
Immerse the sensor module in the control standard solution.
5
Start the measurement with <
C
>.
The calibrated cell constant is displayed.
The unit of the measured value display flashes (AutoRead) as long
as the measured value is not stable.
ESC
TC
Sal
4
mg/l
Lin
Tref20
Tref20
Tref25
nLF
TDS
MEM
C
mS/cm
RCL
S
SAL
O
2
RESET
Cell
LAP
mbar
STO
LoBat Day.Month h:min min:s
Ident
Lap
CLR
M cm
k cm
K
A L
C
CAL
Cell
4 6 8
0
1
cm
CAL
TC
ESC
CellTC
Sal
4
mg/l
Lin
Tref20
Tref20
Tref25
nLF
TDS
MEM
C
mS/cm
RCL
S
SAL
O
2
RESET
Cell
LAP
mbar
STO
LoBat
h:min min:s
Ident
Lap
CLR
M cm
k cm
1
cm
K
A L
C
CAL
Day.Month
1 3 0 5
CAL
TC
<
§_
>
ESC
Day.Month
1
CellTC
Sal
4
mg/l
Lin
Tref20
Tref20
Tref25
nLF
TDS
MEM
mS/cm
RCL
S
SAL
O
2
RESET
Cell
LAP
mbar
STO
LoBat
h:min min:s
Ident
Lap
CLR
M cm
k cm
1
cm
K
A L
C
CAL
2 5 7
CAL
TC
C
<
§_
>