Operation

VARIO Cond

40

Note
If the timer has been stopped you can

l

Reset the timer to the adjusted timer interval with <

R

>.

l

Terminate the timer and switch to the measured value display with
<

E

>.

4.11

Switch-off interval

The automatic switch-off function switches off the meter automatically if no
entry is made for a specified time (switch-off interval). This saves the battery
of the meter.

You can specify the switch-off interval yourself. The switch-off interval is from
10 minutes minimum to 24 hours maximum.

7

If necessary, stop the timer before the time has expired with <

g

>

and make it continue with <

g

>.

8

As soon as the time has expired the alarm signal sounds.
Switch off the alarm signal with <

(

>. The timer goes on counting

with a negative sign.

9

Stop the timer with <

g

>.

1

Switch on the meter with <

(

>.

The display test is displayed.

2

During the display test (approx. 2 sec.):
Display the switch-off interval with <

M

>.

3

Select a switch-off interval with <

>

> and <

<

>.

4

Confirm the setting with <

g

>.

The display switches to the measured value display.
The automatic switch-off function switches off the meter if no entry is
made during the switch-off interval.