6 setting the reference temperature, Setting the reference temperature, Ee section 4.6 s – Xylem VARIO COND TEMPERATURE CONDUCTIVITY METERS User Manual
Page 29: Etting, Reference, Temperature, Refer, Ence
VARIO Cond
Operation
29
4.6
Setting the reference temperature
The setting is only effective if the linear or nonlinear temperature compensa-
tion is switched on.
You can set the reference temperature as follows:
Setting menu of
the reference
temperature
1
Open the setting menu for the cell constant with <
C
>.
The display shows the setting last selected.
2
Open the setting menu for the temperature compensation with <
T
>.
The display shows the setting last selected.
3
If necessary, select the nLF or Lin temperature compensation with
<
§_
>.The display shows the reference temperature set last.
4
Using <
"_
>, switch between tr20 and tr25.
– tr20 = reference temperature 20 °C
– tr25 = reference temperature 25 °C
The adjusted reference temperature is used.
5
Confirm the setting with <
g
>.
– in the nLF menu --> back to the measuring mode
– in the Lin menu --> the menu for setting the temperature coef-
ficient opens up (see section 4.5.2)
<
§_
>
S/cm
CAL
ESC
Lin
Cell
1
cm
0
Sal
mg/l
Tref20
Tref20
TDS
MEM
CAL
C
m
RCL
S
SAL
O
2
TC
RESET
Cell
LAP
mbar
STO
LoBat Day.Month h:min min:s
Ident
Lap
CLR
M cm
k cm
K
r 2
t
n L F
5
Tref25
TC
nLF
Tref25
S/cm
CAL
ESC
Lin
Cell
1
cm
0
Sal
mg/l
TDS
MEM
CAL
C
m
RCL
S
SAL
O
2
TC
RESET
Cell
LAP
mbar
STO
LoBat Day.Month h:min min:s
Ident
Lap
CLR
M cm
k cm
K
r 2
t
n L F
0
TC
nLF
Tref20
Tref20