4.6

Setting the reference temperature

The setting is only effective if the linear or nonlinear temperature compensa-
tion is switched on.

You can set the reference temperature as follows:

Setting menu of

the reference

temperature

1

Open the setting menu for the cell constant with <

C

>.

The display shows the setting last selected.

2

Open the setting menu for the temperature compensation with <

T

>.

The display shows the setting last selected.

3

If necessary, select the nLF or Lin temperature compensation with
<

§_

>.The display shows the reference temperature set last.

4

Using <

"_

>, switch between tr20 and tr25.

tr20 = reference temperature 20 °C

tr25 = reference temperature 25 °C

The adjusted reference temperature is used.

5

Confirm the setting with <

g

>.

– in the nLF menu --> back to the measuring mode

– in the Lin menu --> the menu for setting the temperature coef-

ficient opens up (see section 4.5.2)

<

§_

>

S/cm

CAL

ESC

Lin

Cell

1

cm

0

Sal

mg/l

Tref20

Tref20

TDS

MEM

CAL

C

m

RCL

S

SAL

O

2

TC

RESET

Cell

LAP

mbar

STO

LoBat Day.Month h:min min:s

Ident

Lap

CLR

M cm

k cm

K

r 2

t

n L F

5

Tref25

TC

nLF

Tref25

S/cm

CAL

ESC

Lin

Cell

1

cm

0

Sal

mg/l

TDS

MEM

CAL

C

m

RCL

S

SAL

O

2

TC

RESET

Cell

LAP

mbar

STO

LoBat Day.Month h:min min:s

Ident

Lap

CLR

M cm

k cm

K

r 2

t

n L F

0

TC

nLF

Tref20

Tref20