The Send identification Command responds with sensor vendor, model, and version data. any

measurement data in the sensor’s buffer is not disturbed.

Command

response

“ai!”

“allccccccccmmmmmmvvvxx...xx

where:

a

is the sensor address (“0-9”, “a-Z”, “a-z”, “*”, “?”).

i

is an upper-case aSCii character.

ll

is the SDi-12 version compatibility level, e.g. version 1.2 is represented as “12”.

cccccccc

is an 8 character vendor identification to be specified by the vendor and usually

in the form of a company name or its abbreviation.

mmmmmm is a 6 character field specifying the sensor model number.

vvv

is a 3 character field specifying the sensor version number.

SDI-12 Command & Response

29

where:

a

is the sensor address (“0-9”, “a-Z”, “a-z”, “*”, “?”).

V

is an upper-case aSCii character.

ttt

is a three digit integer (000-999) specifying the maximum time, in seconds, the sensor

will take to complete the command and have data available in its buffer.

n

is a single digit integer (0-9) specifying the number of values that will be placed in the

data buffer. if “n” is zero (0), no data will be available using subsequent “D” commands

example of an “aV!” command:

Command

response

Time

Values

Description

"aV!"

“a0013” 1 sec

3

return fixed data and

diagnostic data for testing

purposes.

Subsequent Command

response

"aD0"

a+123.456+78.9+y

Key

Description

Units

+123.456 Fixed test data

+78.9

Fixed test data

y

rom checksum test

0 = Failed, 1 = Passed

Send identification Command