beautypg.com

Global water – Xylem 6308DT INDUSTRIAL DISSOLVED OXYGEN TRANSMITTERS User Manual

Page 14

background image

Global Water

800-876-1172

globalw.com

- 14 -

(b) 2. 20mA OUT - this value will be used in conjunction with the 4 mA

value to plot the output. (See chapter VII. 4-20 mA OUTPUT.)

8. TEMP. CONTROL SETTING

(a) RELAY 5 - the temperature has only one relay to control you need to

set what action it will use, HIGH or LOW action. . (In HIGH-action
the relay will turn ON if the temperature is greater or equal to
RELAY5 value, in LOW-action the relay will turn OFF if the
temperature is less than or equal to RELAY5 value, which is
modified by the hysteresis value and hysteresis mode.) ((See
chapter VI. CONTROLLING THE RELAYS.)

(b) SET POINT- this is the user changeable value for the Temperature

Alarm relay.