Ebi16 checking the b&d test starting conditions, Place in the chamber, Evaluation – Xylem EBI 16 User Manual
Page 13: Archiving of the results, Positionnement dans la chambre, Archivage des résultats

ba80000def01_EBI16
13
English
Français
EBI16 checking the B&D test
Starting conditions
-
EBI16 must be started at room
temperature (+15°C … +35°C).
Place in the chamber
-
If necessary, preheat the sterilizer
(sterilizer manual)
-
Place the EBI16 in the center of the
chamber.
-
Place the EBI16 vertical in the
chamber.
-
Make sure that the EBI16 is not in the
area of the steam entry.
-
Start the sterilizer immediately after
placement of the EBI16
-
During the B&D test, no other load is
allowed in the chamber.
Evaluation
-
Interface and software are required
for the operation of the EBI16
-
Automatic evaluation of the B&D test
can
only
be
carried
out
with
Winlog.med
or
Winlog.validation.
Archiving of the results
-
The archiving of the results is
ensured permanently by Winlog.med
or Winlog.validation.
Effectuer le test de B&D avec
l’EBI 16
Conditions de depart
-
L’EBI 16 doit être démarré à
température
ambiante
(+15°C…+35°C).
Positionnement dans la chambre
-
Si
nécessaire,
préchauffer
le
stérilisateur
-
Positionner l’EBI 16 au centre de la
chambre
-
Positionner l’EBI 16 verticalement
dans la chambre
-
Veuillez vous assurer que l’EBI 16 ne
se trouve pas à l’endroit où pénètre la
vapeur.
-
Démarrer
le
stérilisateur
immédiatement
après
avoir
positi
onné l’EBI 16
-
Pendant le test de B&D, aucune autre
charge n’est autoriseé dans la
chambre
Evaluation
-
L’interface
et
le
logiciel
sont
nécessaires pour utiliser l’EBI 16
-
L’évaluation automatique du test de
B&D peut être effectuée seulement
avec les logiciels Winlog.med ou
Winlog.validation.
Archivage des résultats
-
L’archivage des résultats est assuré
en
permanence
par
le
logiciel
Winlog.med ou Winlog.validation