Ebi16 checking the vacuum test starting conditions, Place in the chamber, Evaluation – Xylem EBI 16 User Manual
Page 11: Archiving of the results, Positionnement dans la chamber, Archivage des résultats

ba80000def01_EBI16
11
English
Français
EBI16 checking the Vacuum
test
Starting conditions
-
The start temperature is not important
for vacuum test.
Place in the chamber
-
Place the EBI16 in the center of the
Chamber.
-
Place the EBI16 vertical in the
chamber.
-
Make sure that the EBI16 is not in the
area of the steam entry.
-
Start the sterilizer immediately after
placement of the EBI16
-
During the vacuum test, no other load
is allowed in the chamber.
Evaluation
-
Interface and software are required
for the operation of the EBI16
-
Automatic evaluation of the vacuum
test can only be carried out with
Winlog.med
or
Winlog.validation.
Archiving of the results
-
The archiving of the results is
ensured permanently by Winlog.med
or Winlog.validation.
Effectuer le test du vide avec
l’EBI 16
Conditions de depart
-
La temperature de départ n’est pas
importante pour le test du vide.
Positionnement dans la chamber
-
Positionner
l’EBI 16 au centre de la
chambre
-
Positionner l’EBI 16 verticalement
dans la chambre
-
Veuillez vous assurer que l’EBI 16 ne
se trouve pas à l’endroit où pénètre la
vapeur
-
Démarrer
le
stérilisateur
immédiatement
après
avoir
positionné l’EBI 16
-
Pendant le test du vide, aucune autre
charge n’est autorisée dans la
chambre
Evaluation
-
Une interface et un logiciel sont
nécessaires pour utiliser l’EBI 16
-
L’évaluation automatique du test du
vide peut être seulement effectuée
avec le logiciel Winlog.med ou
Winlog.validation
Archivage des résultats
-
L’archivage des résultats est assuré
en permanence par les logiciels
Winlog.med ou Winlog.validation