EBI16 checking the Vacuum
test

Starting conditions

-

The start temperature is not important
for vacuum test.

Place in the chamber

-

Place the EBI16 in the center of the
Chamber.

-

Place the EBI16 vertical in the
chamber.

-

Make sure that the EBI16 is not in the
area of the steam entry.

-

Start the sterilizer immediately after
placement of the EBI16

-

During the vacuum test, no other load
is allowed in the chamber.

Evaluation

-

Interface and software are required
for the operation of the EBI16

-

Automatic evaluation of the vacuum
test can only be carried out with
Winlog.med

or

Winlog.validation.

Archiving of the results

-

The archiving of the results is
ensured permanently by Winlog.med
or Winlog.validation.

Effectuer le test du vide avec
l’EBI 16

Conditions de depart

-

La temperature de départ n’est pas
importante pour le test du vide.

Positionnement dans la chamber

-

Positionner

l’EBI 16 au centre de la

chambre

-

Positionner l’EBI 16 verticalement
dans la chambre

-

Veuillez vous assurer que l’EBI 16 ne
se trouve pas à l’endroit où pénètre la
vapeur

-

Démarrer

le

stérilisateur

immédiatement

après

avoir

positionné l’EBI 16

-

Pendant le test du vide, aucune autre
charge n’est autorisée dans la
chambre

Evaluation

-

Une interface et un logiciel sont
nécessaires pour utiliser l’EBI 16

-

L’évaluation automatique du test du
vide peut être seulement effectuée
avec le logiciel Winlog.med ou
Winlog.validation

Archivage des résultats

-

L’archivage des résultats est assuré
en permanence par les logiciels
Winlog.med ou Winlog.validation