Setup & operation – Xylem H-3553 User Manual
Page 16

SETUP & OPERATION
14
Default Setup
The H-3553 has many settings that can be change. However, the default setups will normally cover most
applications. Table 4-3 shows the default settings for the H-3553.
setting
Default setting
setting range
SDI-12 Address
0
0-9 (Standard), A-Z, a-z
Stage Units
Feet (Slope = 2.3067) Ft, In, M, mm, Ft Dn, Usr Def
Stage Offset
0.00
N/A
SDI-12 Stage Digits 3
N/A
RS-232 Stage Digits 2
Averaging Time
2 Seconds
1-65535 seconds
Bubble Rate
60 bubbles/min
30-120 bubbles/min
Purge Pressure
40 PSI (15 PSI Sensor)
50 PSI (30 PSI Sensor)
30-90 PSI
Purge Sustain
20 seconds
10-40 seconds
4-20mA Min Stage
4.0
N/A
4-20mA Max Stage 20.0
N/A
Table 4-3: H-3553 Default Setup
Reset to Defaults
It may be necessary to reset the H-3553 settings back to factory defaults. Using the RS-232 menu, press the ‘P’
key to enter the “ Advanced Options” menu and then press the ‘D’ key to “ Reset Defaults”.
To reset to defaults using the SDI-12 interface, send the “aXSDEF!” SDI-12 extended command. The response
should be “a0041” which means that it will take 4 seconds to reset to defaults. Note in Table 4-4, the ‘a’ is the
current address of the H-3553.
H-3553 Combo Bubbler setup menu
sDI-12 Interface
P - Advanced Options
Command: aXSDEF!
:
Response: a0041
advanced Options menu
D - Reset Defaults
Resetting to Defaults...
Table 4-4: Reset H-3553 to Factory Defaults