SETUP & OPERATION

14

Default Setup

The H-3553 has many settings that can be change. However, the default setups will normally cover most
applications. Table 4-3 shows the default settings for the H-3553.

setting

Default setting

setting range

SDI-12 Address

0

0-9 (Standard), A-Z, a-z

Stage Units

Feet (Slope = 2.3067) Ft, In, M, mm, Ft Dn, Usr Def

Stage Offset

0.00

N/A

SDI-12 Stage Digits 3

N/A

RS-232 Stage Digits 2
Averaging Time

2 Seconds

1-65535 seconds

Bubble Rate

60 bubbles/min

30-120 bubbles/min

Purge Pressure

40 PSI (15 PSI Sensor)
50 PSI (30 PSI Sensor)

30-90 PSI

Purge Sustain

20 seconds

10-40 seconds

4-20mA Min Stage

4.0

N/A

4-20mA Max Stage 20.0

N/A

Table 4-3: H-3553 Default Setup

Reset to Defaults

It may be necessary to reset the H-3553 settings back to factory defaults. Using the RS-232 menu, press the ‘P’
key to enter the “ Advanced Options” menu and then press the ‘D’ key to “ Reset Defaults”.

To reset to defaults using the SDI-12 interface, send the “aXSDEF!” SDI-12 extended command. The response
should be “a0041” which means that it will take 4 seconds to reset to defaults. Note in Table 4-4, the ‘a’ is the
current address of the H-3553.

H-3553 Combo Bubbler setup menu

sDI-12 Interface

P - Advanced Options

Command: aXSDEF!

:

Response: a0041

advanced Options menu

D - Reset Defaults
Resetting to Defaults...

Table 4-4: Reset H-3553 to Factory Defaults