7.4.1 Resetting the measurement settings
The following settings for pH measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it, press the
key in the measured value display.
7.4.2 Resetting the system settings
The following system settings can be reset to the default condition:
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer
YSI
Cal. interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Resolution pH
0.001
Resolution mV
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
One point calibration
No
Stability control
On
Temperature unit
°C
Setting
Default settings
Language
English
Beep
On
Baud rate
4800 Baud
Output format
ASCII
Contrast
50 %
Illumination
Auto
Switchoff time
1 h