YSI TruLab 1310P User Manual
Page 29

TruLab pH 1310
pH value
ba76156e01
03/2014
29
Continuing with two-
point calibration
6.
Immerse the electrode in the first buffer solution.
7.
When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
8.
Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
9.
Wait for the end of the measurement with stability control or accept the
calibration value with
The calibration display for the setting of the nominal buffer value appears.
10. Set the nominal buffer value for the measured temperature with
<><>.
11. Accept the set calibration value with
The calibration display for the next buffer appears (voltage display).
12. If necessary, finish the calibration procedure as a single-point calibration
with
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.
pH
0.0
24.8 °C
ConCal
7.000
01.02.2014
08:00
Buffer
mV
AR
pH
7.000
24.8 °C
ConCal
1
01.02.2014
08:00
Buffer
13. Thoroughly rinse the electrode with deionized water.