5 calibration points, Calibration points – YSI TruLab 1310P User Manual
Page 31

TruLab pH 1310
pH value
ba76156e01
03/2014
31
5.2.5 Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
26. Set the nominal buffer value for the measured temperature with
<><>.
27. Accept the set calibration value with
The calibration display for the next buffer appears (voltage display).
28. Use
Continue calibrating using the next buffer with
After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
Determined
values
Displayed calibration data
1-point
Asy
Zero point = Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
Zero point = Asy
Slope = Slp.
3-point to
5-point
Asy
Slp.
Zero point = Asy
Slope = Slp.
The calibration line is calculated by lin-
ear regression.
You can display the slope in the units, mV/pH or % (see section
7.1.1 C
pH
9.958
24.8 °C
ConCal
3
01.02.2014
08:00
Buffer