7 settings, 1 measurement settings, 1 changing the settings for ph measurements – YSI TruLab 1310P User Manual

Page 38: Settings, Measurement settings, Changing the settings for ph measurements, Section 7.1.1 c, Hanging, Measurements, Selected (see section 7.1.1 c

background image

Settings

TruLab pH 1310

38

ba76156e01

03/2014

7

Settings

7.1

Measurement settings

7.1.1 Changing the settings for pH measurements

The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required parame-
ter in the measured value display and press the /[menu] or

ENTER> key. After completing the settings, switch to the measured value dis-
play with .
Default settings are printed in bold.

Menu item

Possible set-
ting

Description

Calibration /Cali-
bration record

-

Displays the calibration record of
the last calibration.

Calibration /
Calibration data
storage
/Display

-

Displays the calibration record.

Further options:

Scroll through the calibration
records with <><>.

Output the displayed calibration
record to the interface with
/[USB output].

Output all calibration records to
the interface with [USB
output]
.

Quit the display with /
[Back] or

.

Switch directly to the measured
value display with .

Calibration / Cali-
bration data stor-
age
/ Output to USB

-

Outputs the calibration records to
the interface.

Calibration / Serial
number (probe)

Entry of the serial number of the
connected sensor.
The serial number is output in the
calibration record.

Change the content of the high-
lighted position with <><>.

Go to the next position with
/[].

When the serial number has
been completely entered, con-
firm with .