YSI TruLab 1310P User Manual
Page 30

pH value
TruLab pH 1310
30
ba76156e01
03/2014
Continuing with
three- to five-point
calibration
14. Immerse the electrode in the second buffer solution.
15. When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
16. Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
17. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
The calibration display for the setting of the nominal buffer value appears.
18. Set the nominal buffer value for the measured temperature with
<><>.
19. Accept the set calibration value with
The calibration display for the next buffer appears (voltage display).
20. Finish the calibration procedure as a two-point calibration with
The calibration record is displayed.
pH
4.035
24.8 °C
ConCal
2
01.02.2014
08:00
Buffer
21. Thoroughly rinse the electrode with deionized water.
22. Immerse the electrode in the next buffer solution.
23. When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
24. Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
25. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
The calibration display for the setting of the nominal buffer value appears.